![Copper mountain Planar Tr1300/1 Скачать руководство пользователя страница 30](http://html1.mh-extra.com/html/copper-mountain/planar-tr1300-1/planar-tr1300-1_programming-manual_2666184030.webp)
Planar
TR1300/1,
TR5048,
TR7530
Programming
Manual
SCPI
30
CALC:DATA:SDAT?
CALCulate<Ch>[:SELected]:DATA:SDATa?
Description
Reads
out
the
corrected
data
array
(query
only).
The
corrected
data
array
is
the
data,
whose
processing
is
completed
excluding
the
formatting
as
the
last
step.
Such
data
represent
S–
parameter
complex
values.
The
array
size
is
2N,
where
N
is
the
number
of
measurement
points.
For
the
n–th
point,
where
n
from
1
to
N:
<numeric
2n–1>
the
real
part
of
corrected
measurement
;
<numeric
2n>
the
imaginary
part
of
corrected
measurement.
Target
The
active
trace
of
channel
<Ch>,
<Ch>={[1]|2|3|4|5|6|7|8|9}
Query
Response
<numeric
1>,
<numeric
2>,
…<numeric
2N>
The
data
transfer
format
depends
on
the
FORM:DATA
command
setting.
Related
Commands
FORM:DATA
Equivalent
Softkeys
None
Содержание Planar Tr1300/1
Страница 1: ...Planar TR1300 1 TR5048 and TR7530 Vector Network Analyzer Programming Manual SCPI Oct 2015...
Страница 244: ...244...