232
TRIG:SING
TRIGger[:SEQuence]:SINGle
Description
Generates
a
trigger,
independently
of
the
trigger
source
setting
(except
for
the
External
).
If
the
trigger
source
is
set
to
External
,
an
error
occurs
(error
code
221)
and
the
command
is
ignored.
If
the
instrument
is
not
in
the
waiting
for
a
trigger
state
(sweep
is
in
progress
or
all
the
channels
are
set
to
Hold
),
an
error
occurs
(error
code
211)
and
the
command
is
ignored.
The
command
is
not
completed
before
the
end
of
the
sweep
(waiting
for
the
completion
of
the
sweep
of
all
the
channels).
(no
query)
Related
Commands
TRIG:SOUR
INIT:CONT
INIT:IMM
Equivalent
Softkeys
None