![Copper mountain Planar Tr1300/1 Скачать руководство пользователя страница 24](http://html1.mh-extra.com/html/copper-mountain/planar-tr1300-1/planar-tr1300-1_programming-manual_2666184024.webp)
Planar
TR1300/1,
TR5048,
TR7530
Programming
Manual
SCPI
24
*TRG
*TRG
Description
If
the
trigger
source
is
set
to
LAN
(set
to
BUS
with
the
:TRIG:SOUR
command),
triggers
a
sweep.
If
the
trigger
source
is
not
set
to
the
bus
or
the
instrument
is
not
waiting
for
a
trigger,
the
command
is
ignored.
The
command
is
completed
before
the
end
of
the
sweep.
(no
query)
Target
Instrument
Related
Commands
TRIG:SOUR
Equivalent
Softkeys
None
*WAI
*WAI
Description
Waits
for
the
execution
of
all
commands
sent
before
this
command.
(no
query)
Target
Instrument
Equivalent
Softkeys
None
Содержание Planar Tr1300/1
Страница 1: ...Planar TR1300 1 TR5048 and TR7530 Vector Network Analyzer Programming Manual SCPI Oct 2015...
Страница 244: ...244...