Vascular Use: Vascular System
Product Description

The MedaSonics Tria Vascular Doppler System consists of a speaker unit and
5 MHz Probe.  The probe is connected to the speaker unit via a coiled cord.

Speaker Unit 

The speaker unit is comprised of the electronics, battery compartment,
on/off power control, volume control, on/off and battery life LED indicators.
The 3.5 mm audio output jack on the top of the unit allows private listening
with an optional stethoscope or the connection of a recording device.   

The green LED on the face of the unit is lit when the unit is turned on. When
the battery falls below optimum performance levels, the yellow LED becomes
lit prompting the user to change the battery.

Vascular Probe

The 3/8-inch diameter ultrasound transducer, located at the tip of the probe,
generates signals which are processed and can be heard through the speaker
or optional stethoscope.  The transducer frequency is 5.1 MHz.

Probe Retention

The Doppler probe is held fast to the speaker unit with magnets which can
be seen in the pocket on the right side of the unit.  Simply place the probe
gently in the pocket and the magnets will retain the probe for storage and
transport.  While the retention force of the magnets is high, it cannot prevent
the dislodging of the probe under extreme conditions, such as high impact.

Probe Connection

To remove the probe from the system, gently pull the gray connector out from the
base of the probe. To connect a probe, line up the tab on the end of the connector
with the notch on the end of the wand and push gently until they snap together.

- 10 -

- 9 -

Note: 

The probe on the dedicated vascular Tria System is permanently wired to

the speaker unit and can not be removed.

Battery

The Doppler requires one 9V alkaline battery which is located in a compartment
in the back of the speaker unit. Battery life will vary according to frequency and
duration of exam.

Note: 

A low battery can significantly degrade Doppler performance.  When the

battery falls below optimum performance levels, the yellow LED on the front of
the unit becomes lit and the speaker volume will automatically drop to an unusable
level. This prompts the user to change the battery, returning the Doppler to peak
performance. 

Operation (Speaker Unit)

On/Off Button: 

The On/Off button is located on the left-hand side of the speaker

unit. The green LED on the face of the unit is lit when the unit is turned on.

Volume:

The volume control is located on the left-hand side of the unit below

the on/off switch. Rotate the thumb wheel to achieve the volume that is best for
the application.

Outputs:

The 3.5 mm audio output jack on the top of the unit allows connection

of a stethoscope or recording device.The speaker will automatically mute when
the stethoscope is plugged in.

Instructions for Use

Spread a generous amount of ultrasound coupling agent (ultrasound gel) on
the body surface to be examined. A fluid coupling agent is essential for proper
operation. Lack of signal can be due to loss of fluid coupling. Commercial
preparations provide good coupling, however, a soap solution or even water
can be used. Do not immerse the probe or use alcohol as a coupling agent. 

Содержание Medasonics TRIA P338

Страница 1: ...General Doppler Cautions and Considerations 12 General Doppler Care and Service 13 Battery Replacement 13 Return Policy 13 Warranty 15 Specifications 16 Ultrasonic Output 17 Definition of Terms 18 Tr...

Страница 2: ...Display and 2 MHz Fetal 5 MHz Vascular Probes T345 TRIA Fetal Doppler 2 MHz T346 TRIA Vascular Doppler 5 MHz T336 TRIA Doppler Waterproof Fetal Probe 2 MHz P338 TRIA 2 MHz Fetal Probe P339 TRIA 5 MHz...

Страница 3: ...al heart rate display and one or more Doppler probe s that contain the ultrasound transducer and Doppler electronics Choice of a probe depends on the clinical application The MedaSonics Tria 2 MHz Fet...

Страница 4: ...e level on the Doppler This increases the audio signal upon which the algorithm can operate and will allow it to lock on to the signal and accurately display the heart rate Standard Fetal Probe The 1...

Страница 5: ...ed for proper coupling during underwater use 3 Ensure that the speaker unit and Doppler probe are securely connected through the cord Holding the Doppler probe place the tip on the site to be examined...

Страница 6: ...he end of the connector with the notch on the end of the wand and push gently until they snap together 10 9 Note The probe on the dedicated vascular Tria System is permanently wired to the speaker uni...

Страница 7: ...e before and after such procedures 7 Useful in emergency medicine and noisy environments to take blood pres sure pulse rates and to evaluate suspected arterial injuries 8 Urological applications inclu...

Страница 8: ...s product and can only be performed by your service agent For service contact MedaSonics Service Department at 800 227 8076 or 203 929 6321 Prepaid insured shipment for factory service should be sent...

Страница 9: ...d to weather theft or abuse This warranty does not apply to batteries shipping case or other accessories or to damage from shipping tampering misuse or negligence Preventative maintenance the refinish...

Страница 10: ...e location of the axial spatial peak in cm 18 17 Definition of Terms Ultrasonic Output Tables Note The uncertainty in these measurements was calculated per NEMA UD 2 Appendix A to be 24 ISPTA0 3 The d...

Страница 11: ...e and speaker unit Volume control set Turn up volume too low On LED does not light Battery dead Replace battery when switched on If the problem persists after trying the above suggestions please conta...

Страница 12: ...1...

Страница 13: ...MEDASONICS User s Manual 15 Forest Parkway Shelton CT 06484 800 227 8076 203 929 6321 fax 203 926 9435 www coopersurgical com Part 054 0223 010 Rev C 10 01 Handheld Doppler...

Отзывы: