TRIA System Contents

Thank you for purchasing a MedaSonics Tria

Doppler System. MedaSonics

handheld Dopplers have been a standard of care for 30 years. We are committed
to continuous research and development, marketing and manufacturing of
quality products.

The Tria Doppler System is available with a 2 MHz Fetal Probe, a 5 MHz
Vascular Probe or a 2 MHz Waterproof Fetal Probe. Non-waterproof Doppler
probes can be purchased separately to allow greater versatility in multiple 
clinical settings.

Before using your Doppler, ensure that the carton contains the items shown
below.

1 ea - Speaker Unit or

1 ea - Ultrasound Gel

Speaker Unit w/Display 1 ea - Carrying Case

1 ea - Doppler Probe

1 ea - Warranty Card

1 ea - 9V Battery

1 ea - User’s Manual

If any item is missing or appears damaged or if for any reason you are not
satisfied with your Doppler, please contact your supplier or MedaSonics
immediately.  For more information on accessories or Dopplers, call our
Customer Service Department  at  (800)-227-8076 or (203)-929-6321 (in
the U.S.).  

Introduction

The MedaSonics Tria Doppler System consists of a hand held unit contain-
ing the system controls and speaker, optional heart rate display, and one or
more Doppler probe(s) that contain the ultrasound transducer and Doppler
electronics. Choice of a probe depends on the clinical application.

The MedaSonics Tria 2 MHz Fetal Probe is designed specifically for detection
of the fetal heart rate and circulation. Clinical experience has demonstrated
that Doppler ultrasound is a simple, noninvasive aid in the management of
routine as well as problem pregnancies. 

The MedaSonics Tria 5 MHz Vascular Probe is designed specifically for
detecting blood flow in the arterial and deep venous system of the extremi-
ties.  Clinical experience has demonstrated that Doppler ultrasound is a sim-
ple, noninvasive aid in the assessment of venous and arterial flow in the
extremities.

The MedaSonics Tria Doppler Systems are based on the Doppler shift prin-
ciple. The tip of the Doppler contains two piezoelectric crystals. One crystal
transmits very low power ultrasonic waves at a frequency of approximately
2.3 MHz for the fetal and 5.1 MHz for vascular. As these ultrasonic waves
reach moving tissue such as blood cells, they are reflected and their pitch is
shifted slightly. This change in pitch is the Doppler shift. The receiving crys-
tal detects and amplifies these reflected waves. Further processing converts
the Doppler-shifted ultrasonic waves into the audible sounds of blood flow.  

Содержание Medasonics TRIA P338

Страница 1: ...General Doppler Cautions and Considerations 12 General Doppler Care and Service 13 Battery Replacement 13 Return Policy 13 Warranty 15 Specifications 16 Ultrasonic Output 17 Definition of Terms 18 Tr...

Страница 2: ...Display and 2 MHz Fetal 5 MHz Vascular Probes T345 TRIA Fetal Doppler 2 MHz T346 TRIA Vascular Doppler 5 MHz T336 TRIA Doppler Waterproof Fetal Probe 2 MHz P338 TRIA 2 MHz Fetal Probe P339 TRIA 5 MHz...

Страница 3: ...al heart rate display and one or more Doppler probe s that contain the ultrasound transducer and Doppler electronics Choice of a probe depends on the clinical application The MedaSonics Tria 2 MHz Fet...

Страница 4: ...e level on the Doppler This increases the audio signal upon which the algorithm can operate and will allow it to lock on to the signal and accurately display the heart rate Standard Fetal Probe The 1...

Страница 5: ...ed for proper coupling during underwater use 3 Ensure that the speaker unit and Doppler probe are securely connected through the cord Holding the Doppler probe place the tip on the site to be examined...

Страница 6: ...he end of the connector with the notch on the end of the wand and push gently until they snap together 10 9 Note The probe on the dedicated vascular Tria System is permanently wired to the speaker uni...

Страница 7: ...e before and after such procedures 7 Useful in emergency medicine and noisy environments to take blood pres sure pulse rates and to evaluate suspected arterial injuries 8 Urological applications inclu...

Страница 8: ...s product and can only be performed by your service agent For service contact MedaSonics Service Department at 800 227 8076 or 203 929 6321 Prepaid insured shipment for factory service should be sent...

Страница 9: ...d to weather theft or abuse This warranty does not apply to batteries shipping case or other accessories or to damage from shipping tampering misuse or negligence Preventative maintenance the refinish...

Страница 10: ...e location of the axial spatial peak in cm 18 17 Definition of Terms Ultrasonic Output Tables Note The uncertainty in these measurements was calculated per NEMA UD 2 Appendix A to be 24 ISPTA0 3 The d...

Страница 11: ...e and speaker unit Volume control set Turn up volume too low On LED does not light Battery dead Replace battery when switched on If the problem persists after trying the above suggestions please conta...

Страница 12: ...1...

Страница 13: ...MEDASONICS User s Manual 15 Forest Parkway Shelton CT 06484 800 227 8076 203 929 6321 fax 203 926 9435 www coopersurgical com Part 054 0223 010 Rev C 10 01 Handheld Doppler...

Отзывы: