Guidance and Manufacturer’s Declaration – Electromagnetic Immunity
The CooperSurgical LEEP PRECISION Smoke Evacuation Systems are
intended for use in the electromagnetic environment specified below. The
customer or the end user of the CooperSurgical LEEP PRECISION Smoke
Evacuation System should assure that it is used in such an environment.
Immunity Test
IEC 60601
Test Level
Compliance
Level
Electromagnetic
Environmental -
Guidance
Electromagnetic
discharge (ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Floors should be wood,
concrete, or ceramic tile.
If floors are covered with
synthetic material, the
relative humidity should
be at least 30 percent.
Electrical fast
transient/burst
IEC 61000-4-4
±2 kV for power
supply lines
±1 kV for input/output
lines
±2 kV for power
supply lines
±1 kV for input/
output lines
Mains power quality
should be that of a typical
commercial or hospital
environment.
Surge
IEC 61000-4-5
±1 kV differential mode
±2 kV common mode
±1 kV differential
mode
±2 kV common
mode
Mains power quality
should be that of a typical
commercial or hospital
environment.
Voltage dips, short
interruptions and
voltage variations on
power supply input
lines
IEC 61000-4-11
<5% U
T
(>95% dip in U
T
)
for 0.5 cycle
40% U
T
(60% dip in U
T
)
for 5 cycles
70% U
T
(30% dip in U
T
)
for 25 cycles
<5% U
T
(>95% dip in U
T
)
for 5 sec
<5% U
T
(>95% dip in U
T
)
for 0.5 cycle
40% U
T
(60% dip in U
T
)
for 5 cycles
70% U
T
(30% dip in U
T
)
for 25 cycles
<5% U
T
(>95% dip in U
T
)
for 5 sec
Mains power quality
should be that of a typical
commercial or hospital
environment.
If the user of the
CooperSurgical LEEP
PRECISION Smoke
Evacuation System
requires continued
operation during power
mains interruptions, it is
recommended that the
CooperSurgical LEEP
PRECISION Smoke
Evacuation System be
powered from an
uninterruptible power
supply or a battery.
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency
magnetic fields should
be at levels characteristic
of a typical location in a
typical commercial or
hospital environment.
NOTE
U
T
is the a.c. mains voltage prior to application of the test level; in this case, 230 V.
LP-30-220 • LEEP PRECISION
TM
Smoke Evacuation System, 230 VAC
Operating Manual • English (continued)