1.0 INTRODUCTION 

The DFI 9000 Portable Indicator is a battery-powered signal conditioner and readout device.  The DFI provides 
precise electrical drive to a transducer and converts the transducer output into a digital of engineering units. 
 
The signal conditioner supports transducer outputs between 1 and 4 millivolts/volt (mV/V); a high input option 
supports outputs up to 40 mV/V.  The readout device is a 4-1/2 digit, liquid-crystal display. 
 
Enclosed in a rugged, weatherproof aluminum case, the DFI 9000 is specially designed for remote field operations.  
Quick field setup allows the DFI 9000 to be used for a variety of applications, from checking the load on an oil well 
drilling bit to determining natural gas line pressure. 
 
Coarse zero and Course span controls allow broad adjustment for a wide variety of transducer outputs.  To prevent 
accidental misadjustment, these controls are located on the edge of the circuit board inside the unit, easily 
accessible when the panel is removed from the case.  Fine zero and fine span controls on the front panel allow 
precise adjustment for offsets and gain variations up to +/-15 percent of full scale. 
 
An optional peak detector toggles between peak mode and track mode via a front-panel switch.  In both modes, the 
unit monitors the input signal.  In peak mode, the DFI 9000 stores, digitizes, and displays the highest signal level. 
 
 

2.0 SETUP AND USE 

2.1 Your DFI 9000 

The DFI 9000 can be purchased in one of two ways: 

• 

As a stand-alone unit with a mating connector for the input connector 

• 

As a set with a transducer and a mating connector for the input connector 

 
Some transducers are equipped with internal cables that are used to connect the transducer to the DFI 9000.  
Cooper offers separate cables for transducers without integral cables.  
 
If you purchased Assembly /Calibration with the DFI 9000 and a Cooper transducer, as a system, they are 
calibrated together as a system.   If Assembly/Calibration was not purchased, follow the setup procedures in this 
section before using the unit. 
 
If you purchased the DFI 9000 as a stand-alone unit or if you replace the original transducer, follow the setup and 
calibration procedures before using the unit for the first time. 
 

2.2 Additional Tools 

Use the following tools to set up and calibrate your DFI 9000: 

• 

Two inches of #20 AWG wire (if needed) 

• 

Needle-nose pliers 

• 

Small, standard screwdriver 

• 

Calculator 

• 

Solder and soldering iron (if needed) 

 

2.3 Setup Overview 

1.  Connect the transducer to the unit  
2.  Turn on the unit  
3.  Set the decimal point location  
4.  Verify the calibration  
5.  Fine tune  
 

CF12       DFI 9000  

Rev. C May 2004 

Содержание DFI 9000

Страница 1: ...DFI 9000 PORTABLE DIGITAL INDICATOR USER S GUIDE www cooperinstruments com PH 540 349 4746 FAX 540 347 4755 ...

Страница 2: ...ning 4 3 5 Disable Scaling 4 3 6 Adjust for No Load 5 3 7 Adjust for a Known Load 5 3 7 1 Standard Calibration 5 3 7 2 Shunt Calibration 5 4 0 OPTIONS 6 4 1 Peak Detector 7 4 2 Analog Voltage Output 7 4 3 High Input 7 5 0 MAINTENANCE AND REPAIR 7 5 1 General Care 7 5 1 1 Cleaning 7 5 1 2 Storing 7 5 2 Troubleshooting 7 5 2 1 Low Battery 7 5 2 2 Overrange indications 7 5 2 3 Erratic Readings 8 6 0 ...

Страница 3: ...nd track mode via a front panel switch In both modes the unit monitors the input signal In peak mode the DFI 9000 stores digitizes and displays the highest signal level 2 0 SETUP AND USE 2 1 Your DFI 9000 The DFI 9000 can be purchased in one of two ways As a stand alone unit with a mating connector for the input connector As a set with a transducer and a mating connector for the input connector So...

Страница 4: ...than 30 feet 2 4 2 Connection Use the transducer s integral cable if available If the transducer is not equipped with an integral cable wire your own cable as follows Four wire hookup Figure 1 Disable the sense feature by soldering two small wires inside the mating connector to jumper Excitation pin A to Sense pin B Excitation pin C to Sense pin D DFI 9000 MATING CONNECTOR Six wire hookup Figure 2...

Страница 5: ...all four captive fasteners on the front panel and lifting the panel out of the case 2 Find the decimal point jumper locations 3 Use needle nose pliers to remove the existing jumper and place it in the desired terminals TOP OF FRONT PANEL FIGURE 3 DECIMAL POINT JUMPER LOCATIONS 2 7 Verify Calibration For best results verify the calibration daily Note When not using the original Cooper transducer ca...

Страница 6: ...ils about using the peak detector and other options see Section 4 3 0 CALIBRATION 3 1 When to Calibrate Calibrate the DFI 9000 if you are not using the original Cooper transducer and each time you change transducers 3 2 Methods of Calibration Two methods of calibration are available standard and shunt Use standard calibration when you can place a known load on the transducer Use shunt calibration ...

Страница 7: ... x 19999 15999 100 2 Fine span With the known load applied use a small screwdriver to turn the fine span potentiometer until the display reads exactly 19999 or the scaled down value 3 Scaling This control scales the display to the desired units With the known load applied use a small screwdriver to turn the scaling potentiometer until the display reads the desired full scale engineering units foun...

Страница 8: ...lowing formula Shunt cal output x full scale units scaled shunt cal engineering units Calibration factor With no load on the transducer press and hold the shunt cal button Use a small screwdriver to adjust the scaling potentiometer until the display reads the scaled shunt cal engineering units After scaling return the panel to the case and tighten the four captive fasteners If the display does not...

Страница 9: ... 1 800 344 3921 5 0 MAINTENANCE AND REPAIR 5 1 General Care 5 1 1 Cleaning Clean the exterior and front panel of the DFI 9000 with a damp not wet soft cloth 5 1 2 Storing Store the DFI 9000 at temperature of 20 F to 150 F 5 2 Troubleshooting 5 2 1 Low Battery The red low battery light on the front panel indicated battery voltage below 4 6 volts or an incorrectly installed battery Although the unit...

Страница 10: ... product warranty 6 0 SPECIFICATIONS 6 1 Physical Characteristics Size 7in wide x 5 in deep x 5 in high Number of channels 1 Material Deep drawn weatherproof aluminum case Weight 3 5 lbs 6 2 Environmental Temperature storage 20 F to 150 F Temperature operating 32 F to 130 F 6 3 Power Supply Battery type 4 D cells alkaline recommended Battery life 250 hours for alkaline Note To safeguard the circui...

Страница 11: ...igations of Cooper whether based on contract negligence or otherwise Cooper is not responsible for any incidental or consequential loss or damage which might result from a failure of any and all other warranties express or implied including implied warranty of merchantability or fitness for particular purpose Any unauthorized disassembly or attempt to repair voids this warranty Obtaining Service U...

Отзывы: