background image

 
1.  From the transducer calibration data sheet, obtain the following values: 

- Shunt-cal output (or “shunt-cal factor”) in mV/V 
- Calibration factor (or “full-scale output”) in mV/V 
- Full-scale engineering units (or “capacity”) in psig, lbs, etc. 

2. 

Coarse span

-This control alters the gain of the amplifier or readjusts the span when calibrating to a different 

transducer.  Determine the unscaled shunt-cal display value using the following formula: 
 
Shunt-cal output  x  19999 = unscaled shunt-cal display 
Calibration factor 
 

        

 

With no load on the transducer, press and hold the 

shunt cal

 button. 

Use a small screwdriver to adjust the 

fine span

 potentiometer until the display reads the exact unscaled shunt-

cal display value from step 2. 

3. 

Fine span

-With no load on the transducer. Press and hold the shunt cal button.  Use a small screwdriver to 

adjust the fine span potentiometer on the front panel until the display reads the exact unscaled shunt-cal 
display from step 2. 

4. 

Scaling

-This control scales the display to the desired units.  Determine the scaled shunt-cal value using the 

following formula: 

 

Shunt-cal output  x full-scale units = scaled shunt-cal engineering units 

  Calibration 

factor 

 

        

 

 

 

With no load on the transducer, press and hold the 

shunt cal

 button.  Use a small screwdriver to adjust the 

scaling

 

potentiometer until the display reads the scaled shunt-cal engineering units. 
 
After scaling, return the panel to the case and tighten the four captive fasteners.  If the display does not read 0000, 
fine tune the unit again according to the procedure on page 5.

 

 

Shunt Calibration Example

The following values are taken from a typical transducer calibration data sheet: 

• 

shunt-cal output = 1.4683 mV/V 

• 

calibration factor = 2.9999 mV/V 

• 

full-scale units = 25 psig 

 
To calculate the unscaled shunt-cal display value: 

 

unscaled shunt-cal display= shunt-cal output  x 19999 = 1.4683 mv/v  x 19999 = 9788 

 

 

 

                   calibration factor 

              2.9999 mv/v 

 
To calculate the scaled shunt-cal value: 
   Scaled shunt-cal value = shunt-cal output x full-scale units = 1.4683 mv/v  x 25 psig = 12.236 
 

 

                 calibration factor                              2.9999 mv/v 

 

 

 

 

 

 

 

 

 

4.0 OPTIONS 

The DFI 9000 has three options: 

• 

Peak detector 

• 

Analog voltage output 

• 

High input 

 
See below for brief descriptions of the DFI 9000 options.  For more information about the options, contacts 
Cooper’s Sales Department at 1-800-344-3921. 
 

CF12       DFI 9000  

Rev. C May 2004 

Содержание DFI 9000

Страница 1: ...DFI 9000 PORTABLE DIGITAL INDICATOR USER S GUIDE www cooperinstruments com PH 540 349 4746 FAX 540 347 4755 ...

Страница 2: ...ning 4 3 5 Disable Scaling 4 3 6 Adjust for No Load 5 3 7 Adjust for a Known Load 5 3 7 1 Standard Calibration 5 3 7 2 Shunt Calibration 5 4 0 OPTIONS 6 4 1 Peak Detector 7 4 2 Analog Voltage Output 7 4 3 High Input 7 5 0 MAINTENANCE AND REPAIR 7 5 1 General Care 7 5 1 1 Cleaning 7 5 1 2 Storing 7 5 2 Troubleshooting 7 5 2 1 Low Battery 7 5 2 2 Overrange indications 7 5 2 3 Erratic Readings 8 6 0 ...

Страница 3: ...nd track mode via a front panel switch In both modes the unit monitors the input signal In peak mode the DFI 9000 stores digitizes and displays the highest signal level 2 0 SETUP AND USE 2 1 Your DFI 9000 The DFI 9000 can be purchased in one of two ways As a stand alone unit with a mating connector for the input connector As a set with a transducer and a mating connector for the input connector So...

Страница 4: ...than 30 feet 2 4 2 Connection Use the transducer s integral cable if available If the transducer is not equipped with an integral cable wire your own cable as follows Four wire hookup Figure 1 Disable the sense feature by soldering two small wires inside the mating connector to jumper Excitation pin A to Sense pin B Excitation pin C to Sense pin D DFI 9000 MATING CONNECTOR Six wire hookup Figure 2...

Страница 5: ...all four captive fasteners on the front panel and lifting the panel out of the case 2 Find the decimal point jumper locations 3 Use needle nose pliers to remove the existing jumper and place it in the desired terminals TOP OF FRONT PANEL FIGURE 3 DECIMAL POINT JUMPER LOCATIONS 2 7 Verify Calibration For best results verify the calibration daily Note When not using the original Cooper transducer ca...

Страница 6: ...ils about using the peak detector and other options see Section 4 3 0 CALIBRATION 3 1 When to Calibrate Calibrate the DFI 9000 if you are not using the original Cooper transducer and each time you change transducers 3 2 Methods of Calibration Two methods of calibration are available standard and shunt Use standard calibration when you can place a known load on the transducer Use shunt calibration ...

Страница 7: ... x 19999 15999 100 2 Fine span With the known load applied use a small screwdriver to turn the fine span potentiometer until the display reads exactly 19999 or the scaled down value 3 Scaling This control scales the display to the desired units With the known load applied use a small screwdriver to turn the scaling potentiometer until the display reads the desired full scale engineering units foun...

Страница 8: ...lowing formula Shunt cal output x full scale units scaled shunt cal engineering units Calibration factor With no load on the transducer press and hold the shunt cal button Use a small screwdriver to adjust the scaling potentiometer until the display reads the scaled shunt cal engineering units After scaling return the panel to the case and tighten the four captive fasteners If the display does not...

Страница 9: ... 1 800 344 3921 5 0 MAINTENANCE AND REPAIR 5 1 General Care 5 1 1 Cleaning Clean the exterior and front panel of the DFI 9000 with a damp not wet soft cloth 5 1 2 Storing Store the DFI 9000 at temperature of 20 F to 150 F 5 2 Troubleshooting 5 2 1 Low Battery The red low battery light on the front panel indicated battery voltage below 4 6 volts or an incorrectly installed battery Although the unit...

Страница 10: ... product warranty 6 0 SPECIFICATIONS 6 1 Physical Characteristics Size 7in wide x 5 in deep x 5 in high Number of channels 1 Material Deep drawn weatherproof aluminum case Weight 3 5 lbs 6 2 Environmental Temperature storage 20 F to 150 F Temperature operating 32 F to 130 F 6 3 Power Supply Battery type 4 D cells alkaline recommended Battery life 250 hours for alkaline Note To safeguard the circui...

Страница 11: ...igations of Cooper whether based on contract negligence or otherwise Cooper is not responsible for any incidental or consequential loss or damage which might result from a failure of any and all other warranties express or implied including implied warranty of merchantability or fitness for particular purpose Any unauthorized disassembly or attempt to repair voids this warranty Obtaining Service U...

Отзывы: