background image

NSRTW_mk3 

User’s Manual 

 

 

33 

Select Hourly, Daily or Weekly cycle (or No Secondary Cycle to disable it). Then select the threshold to 
apply during the secondary cycle. Finally select when the secondary cycle begins and when it ends. 

The Start and End of the secondary cycle are defined by: 

 

A Start minute and Stop minute for an hourly cycle 

 

A Start hour/minute and Stop hour/minute for a daily cycle 

 

A Start day/hour/minute and Stop day/hour/minute for a weekly cycle 

When the instrument  time is between the Start and Stop  times, the secondary threshold is applied. 
Otherwise, the primary threshold is applied. That secondary threshold can be lower than the primary 
(for instance for a more sensitive setting at night). It can also be higher. In order to disable alarms 
during certain time periods, use a very high threshold, such as 200 dB. 

Note: 

In order for the secondary threshold to be applied, the instrument time must be set. So after a 

reset of the instrument, and before the time is set by running the Instrument Manager, or the instrument 
connecting to a server, the secondary threshold cycle will not be in effect, even if it was programmed. 
This is because any secondary threshold is based on the instrument internal time, and that cannot 
operate properly until the proper time is set. 

7.6.7

 

WiFi™

 Battery Drain 

Connecting to a network drains the battery. Even if the instrument is idle, if the periodic connection is 
enabled, the instrument will regularly wake up and attempt to connect to the server. Trying to connect 
at regular intervals will drain the battery in a short amount of time. 

When that behavior is unwanted (for instance when the instrument is unused and in storage) pressing 
the 

Disable All

  button  (see 4 

Figure 5

)  will completely disable 

WiFi™

  and email  connectivity. If the 

instrument is idle with all its 

WiFi™

 connectivity disabled, it can keep its battery for at least 6 months. 

8

 

Instrument_Listener Application 

In order to report recorded data, and send alarm emails, the instrument must connect to a remote PC 
(a server) that is running the 

Instrument_Listener

 application. The 

Instrument_Listener_Installer_Vxxx

 

must be installed on the PC (

Vxxx

 represents the version number of that application). 

The remote PC (the server) must: 

 

Be on 

 

Be reachable over the network, either by an instrument on the local network, or on a global 
network. In the case of a global network, all firewalls must allow the server to receive incoming 
connections. 

 

Be running the 

Instrument_Listener

  application. It may be useful to setup that application so 

that it runs automatically when the PC boots. 

The 

Instrument_Listener

  application (see 

Figure 13

) always listens to incoming connection attempts 

from instruments in the field. Whenever an instrument tries to connect, it opens an 

NSRTW_Server

 

window (see 

Figure 14

). That window briefly opens to show the communications with the instrument, 

battery level and temperature. It manages all operations that the instrument needs to perform and then 
closes back down.  

Содержание NSRTW mk3

Страница 1: ...NSRTW_mk3 User s Manual January 29 2019 Bruno Paillard ...

Страница 2: ..._____________ 7 6 4 2 Operation LED _____________________________________________________________ 8 6 5 Sound Level Calculation _________________________________________________________ 8 6 6 Manual Recording ______________________________________________________________ 9 6 7 Automatic Recording with WiFi Connection W model only ______________________ 10 6 7 1 WiFi Connectivity __________________...

Страница 3: ...______________________________________________ 25 7 6 2 Cloud Setup _______________________________________________________________ 26 7 6 3 WiFi Settings Without Cloud Service __________________________________________ 29 7 6 4 Manual Connection _________________________________________________________ 30 7 6 5 Setting Up the Server to Work in a Local or Wide Area Network _____________________ 30...

Страница 4: ...r outdoors applications All digital design Ultra stable sensitivity field recalibration is easily done but seldom required Very low sensitivity variation due to temperature changes Very low sensitivity to vibrations Software function calculates global Leq and or dose according to ISO and OSHA methods Adjustable response time Preprogrammed recording start date time Integrated oscilloscope function ...

Страница 5: ... is facing down This way rain will not accumulate in the connector and microphone port 4 Getting Started 4 1 Software Installation On the PC that is used to setup the instrument run Instrument_Manager_Installer exe This installs the Instrument_Manager application as well as its USB driver and documentation On the PC that is used as a server to collect data over the network W model only run Instrum...

Страница 6: ...ument time to be synchronized to the PC time Only say YES if the PC time is correct 5 In addition the application detects if the time zone of the instrument is different from the time zone of the PC If it is the application asks to set the instrument to the same time zone as the PC Note If in rare cases the instrument is not intended to be used in the same time zone as that of the PC used to confi...

Страница 7: ...e increased If the connection fails the NSRTW_mk3 will fall back to its previous state Sleep or Recording Note WiFi communications consume battery so when leaving the NSRTW_mk3 unused for a long time it is important to disable the WiFi periodic connection feature In applications where the NSRTW_mk3 must record for longer periods of time it can be connected to a USB charger https convergenceinstrum...

Страница 8: ...up to 6 months Active Connected to USB Not recording Instrument is fully on and measuring Instrument draws power from USB to power itself and if necessary recharge its battery Battery is not depleted Recording USB connected or not Fully on and recording If USB connected power comes from USB If not power comes from the internal battery If not connected the instrument can record for up to 7 days Per...

Страница 9: ... time constant is applied to the calculation of L min and L max but not Leq A typical short time constant 125 ms allows the measurements to track sharp changes while a longer time constant 1s provides more stable cleaner RMS values that averages out short transients The Leq is integrated linearly over the specified log interval irrespective of the time constant setting The sound level statistics L...

Страница 10: ...memory That occurs typically when a recording is started and stopped quickly for a time shorter than the log interval before any data had time to be written to memory In that case only the time stamp is written to memory An empty record can also occur when a recording was started but no data Lmin Lmax or Leq was selected to be recorded To record sound levels proceed as follows 1 Select the statist...

Страница 11: ...a the timer The instrument will be completely controlled by the server including starting the recording erasing memory if required and resetting the time When using the WiFi feature the instrument will try to connect to a known server at regular intervals or immediately in case of alarm Whenever it connects the instrument will talk to a special Instrument_Listener application running on the server...

Страница 12: ...o alarms can be enabled A noise level over a preset threshold The battery is close to being depleted 6 7 1 3 WiFi Reporting Via Cloud Service Reporting and email alerts via the cloud service work exactly like reporting without the cloud service except that it uses Convergence Instrument s servers and the user does not have to deploy their own server for the instruments to be able to connect In add...

Страница 13: ...ton 3 Memory fill indicator 4 Timer button 5 Record button 6 Instrument Time indicator 7 Tab selector 8 Instrument Temperature indicator 9 Erase button 10 Instrument User ID 11 Instrument Information 12 Settings button 13 Field calibration button 14 Battery Condition Charge and Recording Time Field 15 Sound Level indicator ...

Страница 14: ...evels 3 dB Lin button 4 Pan and zoom buttons 5 Record number selector 6 Cursor 7 Data source 8 File Save button 9 File Open button 10 Instrument Download button 11 Export button 12 Global level dose calculation method 13 Global Level Time Weighted Average value 14 Percentiles ...

Страница 15: ...SRTW_mk3 User s Manual 14 Figure 4 Spectrum Tab 1 Cursor Zoom and Pan controls 2 Cursor values display 3 dB Lin button 4 Auto scale button 5 Cursor 6 Spectrum 7 Spectral controls 8 Spectrum reset button ...

Страница 16: ...int Connected Indicator 11 Network Connected Indicator 12 Server IP address 13 Server Connection Indicator 14 Email Active Indicator 7 1 Starting the Application Note The application can be run with or without an NSRTW_mk3 connected When an NSRTW_mk3 is not connected the application can only be used to display and analyze a previously saved log file All the controls and indicators related to the i...

Страница 17: ...rum Tab Is used to display a real time power spectrum from the NSRTW_mk3 see Figure 4 The WiFi Tab Is used to display the present connection state of the NSRTW_mk3 and to setup its connection parameters SSID Security Password etc see Figure 5 7 3 Setup Tab The setup tab is seen in Figure 2 It is divided into seven fields 7 3 1 Time and Date Field This field displays the internal time of the instru...

Страница 18: ...Timer button is used to set a time and date to start the recording automatically This is only allowed when a recording is not already in progress To set the timer press the Timer button Then set the recording date and time and press the OK button to arm the timer Erase The Erase button erases unconditionally all the data that has been recorded in the instrument making room for new records The Memo...

Страница 19: ...thout changing the date time setting 2 OK button Closes the window and accepts the newly set date time 3 Set to Now button Set the time and date to the present time 4 Date setting field Click a date to set 5 Time setting field Use the field to set a new time 6 Window close button Closes the window without changing the date time setting 7 3 5 Settings Field The settings field contains the Settings ...

Страница 20: ...mates the sensitivity of human hearing 7 3 5 1 2 Bandwidth Selector Use this selector to choose the sampling frequency bandwidth 16 kHz bandwidth 32 kHz sampling frequency 24 kHz bandwidth 48 kHz sampling frequency 7 3 5 1 3 Log Interval The log interval defines the time between two successive recorded points That also defines the integration period for the LEQ and the observation period for L min...

Страница 21: ...ime constant is not used for the calculation of Leq In the case of LEQ the sound energy is averaged linearly over the specified log interval 7 3 5 1 6 Closing the Window Each adjustment is implemented in the instrument when the OK button is pressed If the window is closed without clicking OK then the settings are not implemented 7 3 5 1 7 Save Recall Once the parameters are adjusted the whole setu...

Страница 22: ...ing curve 7 4 Records Tab The Records tab is seen in Figure 3 It is used to retrieve the recorded information from the NSRTW_mk3 or from a saved file and to display and optionally export the recorded logs 7 4 1 Instrument Download To download data from the instrument press the Instrument Download button The whole contents of the instrument s memory are downloaded and the newest record is displayed...

Страница 23: ... the network When a wlg file is opened the Instrument_Manager application asks to convert it to a cil file The Source field shows the User ID of the instrument the data comes from That function works even when no instrument is connected to the PC so it can be used to examine data that was recorded elsewhere and sent to the user as a file 7 4 4 Exporting Data The data of the record being presently ...

Страница 24: ...trol simply press the button that looks like a magnifying glass and choose one of the following zoom functions Zoom X Y Zoom X Zoom Y Zoom Extents After choosing the function the cursor changes to a magnifying glass Click and drag on the graph to apply the required zoom 7 4 7 2 Pan Control To pan a graph left right up or down first press on the Hand button to the left of the graph Then position th...

Страница 25: ...ation A larger value of Nb_Avg produces a spectrum that is more stable but also averages out short lived transients Linear Averaging takes an equal weight average of Nb_Avg successive spectra After the Nb_Avg spectra have been observed the result is frozen In both cases the Reset button discards the present display and restarts the averaging from zero The Averaging Mode can be chosen from several ...

Страница 26: ...rt of the regular reporting of recorded levels The server connection can be done to a specified IP address or to a domain name Connecting using a domain name is preferable because IP addresses are often subject to change The instrument has a total time of 1 minute to go through the two stages If the final stage is not established by the end of that minute the instrument stops trying in order to pr...

Страница 27: ... must be configured activated and deactivated using that Cloud Setup function In addition to simplifying the configuration process it insures that there is no discrepancy between the internal configuration of the instrument and the Cloud database When the instrument is used on the cloud the normal WiFi Setup Email Setup and Disable All functions of the Instrument Manager should not be used In addi...

Страница 28: ... You must adjust all required settings and press the Submit button at the bottom of the web page a Instrument Info Add change the user ID for that instrument if one is required b Instrument Configuration Adjust all the measurement parameters for the particular instrument The setup for a VSEW_mk2 is presented Measurement setup options will be different for another type of instrument For a VSEW_mk2 ...

Страница 29: ...ured The instrument may try to connect immediately or it may try to connect after the connection interval time has elapsed If the instrument is not actively trying to connect right away simply press the Connect Now button 6 Once the instrument has successfully connected to the server the Manager can simply be closed and the instrument disconnected from USB The instrument will keep connecting at le...

Страница 30: ...e instrument report much more often but that will drain the battery faster Setting a much longer interval several days a week or more will risk losing data if the connection cannot be established at its scheduled time The interval must not be set as 0 We do not advise setting a schedule interval lower than 2 min The WiFi Report Active button must be pressed When the button is released the instrume...

Страница 31: ...ove one should make sure of the following The IP address of the PC working as a server must be fixed To that end the router that the PC is connected to must be set to use Static DHCP for that PC This way the IP address cannot change The PC working as a server must be always on and must be running the Instrument_Listener Otherwise the connection can t be established when the instrument tries to con...

Страница 32: ... Gmail_Setup pdf for details The complete email setup is similar to the configuration of any email application It includes the following Email server The internet name of the outgoing email server For gmail this is smtp gmail com Port number The TCP port number of the email service This is usually o 25 for an open no security account o 465 for SSL security o 587 for TLS security For gmail this is ...

Страница 33: ...7 6 6 1 Advanced Email Setup Firmware versions of the instruments from V1 7 and up have the following advanced capabilities 7 6 6 1 1 Email Rate Limits The rate at which the alarm emails are sent can be limited By default it is 1 email per 5 min for a noise alert and 1 email per hour for a battery email These limits can be increased or decreased by writing new values in the respective fields 7 6 6...

Страница 34: ...dic connection is enabled the instrument will regularly wake up and attempt to connect to the server Trying to connect at regular intervals will drain the battery in a short amount of time When that behavior is unwanted for instance when the instrument is unused and in storage pressing the Disable All button see 4 Figure 5 will completely disable WiFi and email connectivity If the instrument is id...

Страница 35: ...dy been created If the instrument is not currently recording it then erases the instrument s memory adjusts the time and starts recording If the instrument is recording it closes as is and lets the recording continue 8 1 File Storage The wlg files that are created by the NSRTW_Server are stored in C Users username Documents Convergence_Instruments All_Instruments Records NSRTW_mk3 where username i...

Страница 36: ...ment However a hardware reset will reset the instrument s time and date to a value in year 1903 9 3 Operation Outdoors The NSRTW_mk3 is weather proof It can safely be used outdoors provided that the temperature is always within the safe limits However when using it outdoors with a risk of rain or snow make sure its USB connector is facing down so water will not accumulate in the connector and micr...

Страница 37: ...st an indication that the port is powered It should light up even if the USB driver is not installed properly 5 Make sure that the operation LED starts blinking If not go to step 6 6 Open the Device Manager on the PC This is usually found in Control Panel System and Maintenance 7 Just after connecting the instrument to the PC observe that the Device Manager window refreshes 8 Verify that an item n...

Отзывы: