background image

WATER PREPARATION & MAINTENANCE

To ensure proper operation, the spa must be drained, rinsed, refilled with fresh water, and balanced to the recom

-

mended levels indicated in this section before installing the product.  Use the test strips to measure water chem

-

istry and verify that the water has been balanced properly.  Replacing the filter is recommended but not required 

unless the spa was previously sanitized with a hydrogen peroxide system.  The combination of hydrogen peroxide 

and chlorine may cause gum-like build up, water discoloration, and skin irritation.

It is important to properly maintain your spa for the health and safety of its users.  Spas are unique in both size 

and usage patterns.  ChlorMaker IL will significantly reduce the amount of spa maintenance required, but chemi

-

cal check-ups including chlorine levels and PH are recommended before each use.  Proper water maintenance 

will prolong the life of ChlorMaker IL and the spa. The following table describes generally accepted optimum water 

chemistry for portable spas.

PARAMETER

RECOMMENDED LEVEL

Free Chlorine

3.0 to 5.0 PPM

PH Balance

7.2 to 7.6

Total Alkalinity

80 to 120 PPM

Calcium Hardness

300 to 400 PPM

It is recommended that chlorine and PH levels are checked before each use, or at least once per week when not in 

use.  Alkalinity, Calcium Hardness, and Salt Concentration should be checked at least once per month.  It is recom-

mended that the spa is manually shocked after high usage (example: after a party with multiple users), or biweekly 

if used frequently.  Always remember to leave the spa circulating for several hours before retesting.  Phosphates in 

the spa water will increase the demand for chlorine and will reduce the life expectancy of the generator.  Remove 

Phosphates upon start up and each month there after.  Phosphate Remover can be purchased at any local Pool & 

Spa supply store.

WARNING:

  

Consistent chlorine levels above 5.0 PPM or consistent salt concentration levels above 5,000 

PPM may lead to corrosion of metal components in the spa.
WARNING:  Consistently low Calcium Hardness levels below 150 PPM may lead to spa equipment damage 

and failure.  Make sure Calcium Hardness levels are within the recommended range above.

 

Содержание ChlorMaker IL

Страница 1: ...ControlOMatic Inc 12659 Arbor Lane Grass Valley CA 95949 www ControlOMatic com Support ControlOMatic com 530 205 4520 Manual Version 2 1 14 Made in the USA ChlorMaker IL TM OPERATING INSTRUCTIONS...

Страница 2: ...otected VAC power source only WARNING Protect the power supply and outlet from exposure to the elements including direct sun rain snow condensation etc WARNING DO NOT use with extension cord Injury ma...

Страница 3: ...er chemistry for portable spas PARAMETER RECOMMENDED LEVEL Free Chlorine 3 0 to 5 0 PPM PH Balance 7 2 to 7 6 Total Alkalinity 80 to 120 PPM Calcium Hardness 300 to 400 PPM It is recommended that chlo...

Страница 4: ...he volume of water in gallons or liters by checking the spa owner s manual or use any internet volume calculator 99 Pure Sodium Chloride NaCl Coarse Grain Example For a 350 Gallon Spa Divide 350 Gal b...

Страница 5: ...control panel to the power source The black cable clips can be used to secure the cable in place IMPORTANT Make sure the power source is GFI Protected and is protected by a weather protec tive cover S...

Страница 6: ...ure below Use the supplied screws to securely fasten the bottom section of the control panel to the skirting 6 Fasten Top Cover of Control Panel Make sure the black rubber gasket is secured in the gro...

Страница 7: ...agram below Item A Cell Orientation The cell must be mounted relatively vertical to prevent chlorine gas from being trapped in the cell 9 Connect Tubing To Cell Cut the existing tube and reconnect the...

Страница 8: ...is important to re set the power level down several levels If usage significantly increases return from vacation etc it is important to adjust the power level up accordingly Always retest and adjust t...

Страница 9: ...ghts Salt level may be too high and engaged the automat ic shut down feature Dilute the spa water by draining some of the spa water and refilling with fresh water Plates Are Dirty With White Or Scale...

Страница 10: ...Hex Nut at top of cell housing counter clockwise with a large wrench as shown in picture below The cell plates will turn with the Hex Nut Loosen the Hex Nut to a point that can be turned by hand Loos...

Страница 11: ...ct or to explain the usage of the system to the buyer repairs necessitated by use other than normal home use damage resulting from misuse unintended use unforeseen use non pool or spa use abuse accide...

Страница 12: ...prod uct Or register your product on line at www ControlOMatic com First Name Last Name Address City State Province Postal Code Phone E Mail Please copy this section and forward a copy to ControlOMat...

Отзывы: