Installation & Maintenance
515 FA01 - 29 August 2021
11
Figure 5 shows the new option card with the optional Ethernet port in place
of RS-485 port (terminals 19-21).
Figure 5 Rear Panel Connections - New Ethernet Version
Terminal Designations
Terminal Label
Designation
Comment
1 FINP
1+ Frequency Input 1+
Channel 1 Flow Input
2 FINP
2+ Frequency Input 2+
Channel 2 Flow Input
3 SG
-
Signal ground
4 EXC V
1+ Excitation Term 1+
Not used
5 EXC V
2+ Excitation Term 2+
Not used
6 EXC V
3+ Excitation Term 3+
Not used
7
AINP1
+ Analog Input ch 1 (+)
Not used
8
-
Analog Input ch 1 (-)
9
AINP2
+ Analog Input ch 2 (+)
Not used
10
-
Analog Input ch 2 (-)
11
AINP3
+ Analog Input ch 3 (+)
Channel 1 Flow Input
12
-
Analog Input ch 3 (-)
13
AINP4
+ Analog Input ch 4 (+)
Channel 2 Flow Input
14
-
Analog Input ch 4 (-)
15 Vo
+ 8-24 volts DC output
Overload protected
16 G
-
DC Ground
17 Vi
+ DC power input
DC power in 12-28V
18 SH
E Shield terminal
E
AC
MAINS
E Mains ground
AC power in 100-
240VAC
N
N Mains neutral
A
A Mains active
19
RS485
COM-2
port
+ RS485 (+)
Optional RS485 port may
be replaced by Ethernet
port.
20
-
RS485 (-)
21
G RS485 ground
22
LOGIC
INPUTS
1+ Switch 1
23
2+ Switch 2
24
3+ Switch 3
Remote Reset
25
4+ Switch 4
CAL Switch – In field
access protection
26
C- Signal ground
27
OUT1
+ Output ch 1 (+)
28
-
Output ch 1 (-)
29
OUT2
+ Output ch 2 (+)
30
-
Output ch 2 (-)
31
RELAYS
RC Relay Common 1-2
Term 31 - Common 1-4
on legacy option card
32
R1 Relay 1
33
R2 Relay 2
34
R3 Relay 3
35
R4 Relay 4
36
RC Relay common 3-4
Term 36 only available on
new style option card
RS232 COM-1 port 9-pin serial port
Terminal Label
Designation
Comment
Содержание 515 FA01
Страница 8: ...viii 515 FA01 29 August 2021...
Страница 10: ...x 515 FA01 29 August 2021...
Страница 86: ...Communications 76 515 FA01 29 August 2021...
Страница 88: ...78 515 FA01 29 August 2021...
Страница 94: ...84 515 FA01 29 August 2021...