Instrument Calibration
60
515 BR03 - 15 August 2021
Time Settings and Data Logging
Instrument Clock
The instrument has a real-time clock for recording logged events. The clock
displays the time and the date. The date format can be set to European
format (day/month/year) or American format (month/day/year). The time
clock uses the 24-hour format.
The clock will continue to operate for up to 5 years (typically) on the
internal battery if there is no power connected to the instrument. Therefore,
after an interruption to the power supply, the instrument recommences
normal operation although there will be no data recorded during the period
without a power supply.
Note:
If there is an interruption to the power supply and the battery has
failed, the instrument displays an error message when the power
supply is restored. In this case, you should set the current time and
date so that the instrument continues to log data at the correct times.
RTU
DATA
The Modbus RTU data format for the 2-register (4-byte) values can be
set as either floating point or long integer values.
Use
or
to select FLOAT or INTEGER.
RTU
ADDR
The Modbus RTU protocol address must be in the range of 1 to 247.
When multiple instruments (slaves) are connected to one communication
device (master), each assigned address must be unique.
Note:
The master device uses the RTU address 0 (zero) for broadcasting
to all connected slave units.
ASCII
ADDR
The ASCII protocol address identifies each communicating device.
The address must be in the range of 1 to 255. When multiple instruments
(slaves) are connected to one computer (master), each assigned address
must be unique.
FLASH
PORT
The Flash Driver Port assignment defines the communication port for
downloading software into the instrument.
The default setting of this assignment is the RS-232 port. The Ethernet
port, if fitted, can not be used as the Flash port.
Press
or
to select RS-232 (COM-1), RS-485 (COM-2) or
NONE.
UNITS PARAMS INPUTS OUPUTS ALARMS
COMMS
TM/LOG SETUP TEST END
SET
Содержание 515 BR03
Страница 8: ...viii 515 BR03 15 August 2021...
Страница 10: ...x 515 BR03 15 August 2021...
Страница 16: ...Introduction 6 515 BR03 15 August 2021...
Страница 78: ...Instrument Calibration 68 515 BR03 15 August 2021...
Страница 100: ...90 515 BR03 15 August 2021...
Страница 106: ...96 515 BR03 15 August 2021...