30
Contour M2 device characteristics
device Name:
Contour Technology M2
Intended Use:
The Contour Technology M2 device is intended for
stimulation of healthy
muscles in order to enhance and facilitate improved
muscle performance.
Waveforms:
SwissWave
™
- The waveform is monophasic
rectangular, voltage regulated.
Parameters:
The parameters (pulse rate, pulse duration, stimulation
time, rest time) are optimized for each program.
Pulse duration:
Maximal impulse duration between 200-500 µs
(depending on program)
Protection against
electric shock:
Internal electrical power source
Maximum Output:
52 Volt +/- 10%
dC Component:
3.2 Volt +/- 10%
Channels
:
2 independent, isolated channels
Separate Intensity control levels for each channel
Power Supply:
4 AAA battery 1.5 V (Alkaline, NiCd or NiMh)
Approx. max consumption:
130 mA at 100% intensity
Approx. normal consumption:
40 mA at 100% intensity
Battery life:
20 hours at 100% intensity
Pulse Frequency Range
1 – 120 Hz
FdA Clearance
Nonprescription, over-the-counter use
The M2 is manufactured in accordance with the requirements of International
Standards IEC 60601-1, IEC 60601-2-10. The M2 conforms to requirements of the
United States Food and Drug Administration (FDA) for non-prescription sales.
BF (body protected) symbol. This means that the outputs are isolated
from other parts of the M2 device that provide a protection against
electrical shock.
This symbol means that the supplied documentation must be consulted
for operating, cautionary or safety information before using the device.
Contour CoreBelt
™
characteristics
Outer Material:
Neoprene 100%
Inner Material:
Terry Cloth
Polyester 85%, 15% Cotton
Belt:
Nylon 100%
Parameter
500
Ω
1000
Ω
1500
Ω
Output RMSV
6.50 V
6.55 V
6.60 V
Output RMSA
13.00 mA
6.55 mA
4.40 mA
Output Frequency
50Hz
50Hz
50Hz
DC Component: Approx.
32 µC
16 µC
11 µC
Positive Pulse Width:
300 µs
300 µs
300 µs
31
Care and cleaning
To clean the Contour M2 device, use a soft cloth lightly dampened with an alcohol-
based cleaning product, that does not contain solvents. Solvents could damage the
plastic parts, especially the display.
Your Contour M2 device is designed to be maintenance free.
There are no user serviceable parts inside.
The Contour CoreBelt
™
is machine washable in warm water/gentle cycle.
You must always follow the instructions on the label when washing the belt
• Do not use any bleach when washing your belt
• Do not dry clean your belt
• Do not tumble dry your belt. You should always dry the belt on
a flat surface
• Do not dry over anything hot as the belt contains plastic parts.
You should ensure the belt is completely dry before use
• Do not iron
Warnings concerning battery handling
Always read and follow the specific instructions of the battery manufacturers.
Note:
• When inserting batteries ensure that battery polarity is correct.
• Do not expose batteries to temperatures exceeding the
manufacturer's specifications.
• Do not store and/or ship this unit with batteries inserted.
• Do not use lithium batteries.
• Do not attempt to recharge alkaline batteries.
• Do not dispose of any batteries in a fire.
• Note that batteries may present burn or fire hazard if short-circuited.
• Improper battery handling may result in explosion, leakage or flames.
• Be sure to remove the batteries when the unit is not used on a
regular basis.
• Please apply regulations enforced in your country when disposing your product
at its end of life. Therefore, please discard it in a location planned to this effect,
ensuring it is disposed in an environment-friendly manner.
This is also applies to accessories.
Electrode disposal
Never dispose Gel Pads (new or used) in a fire. Dispose of in accordance with local
regulations. Please contact your local authorities if you need more information.