Problem

Possible Cause

Solution

I am not feeling 
anything

Gel Pad not directly against 
your clean, dry skin

Make sure the plastic liner is removed from 
your gel pad electrode and place it (BLACK 
SIDE) directly against your clean, dry skin.

Intensity level might not be 
high enough

The intensity must be at least 15% before 
you will feel anything.

Gel Pads may not be 
properly positioned

Turn the M2 device off, remove the Core-
belt

 and reposition as described in this 

manual. Make sure the Gel Pads are below 
your rib cage.

The Lead-Wire module is 
not inserted correctly

Make sure the wires are securely attached

I pressed power 
and the device 
doesn't turn on

The batteries are inserted 
improperly

Replace batteries in proper position as 
shown on the set-up pages

The batteries are dead

Replace all 4 AAA batteries
Never mix old and new batteries

Display is on but 
there is no signal

The device is paused

Press On/Off button

The Lead-Wire module is 
not inserted correctly

Slide the Lead-Wire module onto groves 
of the M2 device until you hear it click 
into place

The Lead-wire(s) may not 
be connected to the Gel 
Pads

Turn the M2 device off and insert the
pin-end of the electronic module wire into 
the wire from the Gel Pad

Gel pad electrodes are 
not making contact with 
your skin

Make sure the black side of the gel pad 
electrode is making good contact wIth your 
clean, dry skin

The battery
symbol is flashing

The batteries are low

Replace all 4 AAA batteries.
Never mix old and new batteries

The contractions 
are weak even 
though the
intensity is high

The batteries are low

Replace all 4 AAA batteries.
Never mix old and new batteries

The Gel Pads have
lost conductivity

Replace Gel Pads

The pads are positioned 
improperly

Turn the M2 device off, remove the Core-
belt

 and reposition as described in this 

manual. Make sure the Gel Pads are below 
your rib cage.

troubleshooting

27

Problem

Possible Cause

Solution

Stimulation is 
uncomfortable

The Gel Pads may
be worn out

Replace Gel Pads

Intensity level might be set 
too high

Reduce the intensity level until your body 
gets used to it.

Gel Pads may not
be properly positioned

Turn the M2 device off, remove the 
Corebelt

 and reposition as described in 

this manual. Make sure the Gel Pads are 
below your rib cage

Too many sessions in a row

Leave at least six hours between
sessions

My muscles are 
sore after using the 
Contour CoreBelt

Intensity level might be set 
too high

Reduce the intensity level

I feel tingling in
my legs

Gel Pads are located near a 
nerve center

Move the Gel Pads slightly

I just bought the 
Contour but I’m not 
seeing results

Your diet plays a vital 
role.  Are you following the 
Contour diet plan?
 

Follow the eating plan in the Diet and 
Exercise Guide

The intensity level might not 
be set high enough

Increase the intensity level.
You should see results in 2-3 weeks 

Содержание CoreBelt

Страница 1: ......

Страница 2: ..._____________10 Intensity The Key to an Effective Workout__________________12 Getting Started With Total Body Exercise___________________14 Training Programs____________________________________16 Scheduling Contour Training Sessions_____________________18 Work Table__________________________________________20 Frequency Table______________________________________21 Frequently Asked Questions_______...

Страница 3: ... and caution with electrodes sections in this manual The various types of muscle work that the Contour CoreBeltimposes on the stimulated muscles are able to improve or facilitate muscle performance in addition to firming tightening and strengthening muscles The Contour CoreBelt can therefore be considered a technique for muscle training How it works The Contour Technology Ab Belt features four spe...

Страница 4: ... increase and decrease intensity Sound Mute Select Information i LCD Display On Off Pause 4 The M2 Electronic Muscular Stimulator The Contour M2 Electronic Muscular Stimulator with technology that mimics the way your brain controls muscles The M2 Electronic Muscle Stimulator display 5 ...

Страница 5: ...the top edge of the belt Repeat steps 5 and 6 for placing remaining Gel Pads along the back of the CoreBelt 7 This side to garment T belt loop left side of the CoreBelt 7 ABS ABS This side to garment T belt loop left side of the CoreBelt 8 ABS ABS This side to garment T belt loop left side of the CoreBelt 9 Slide the two Gel Pad connector wires through the hole on the left side of the CoreBelt Rep...

Страница 6: ...ply push the Select button to advance to the next screen Turn up the intensity 6 TURN ON THE Contour M2 device 1 Push the On Off button A logo screen will appear followed by a warning screen We recommend you read the warning screen and refer to the warning chapter in this manual if you have any concerns Select Recent Program 2 After a few seconds the screen will ask you would you like to use the m...

Страница 7: ...nd you should increase the intensity to achieve the best results Good Better Best Intensity Levels Pressing the On Off Button during the exercise phase will pause the program To continue simply press the toggles Pressing the On Off Button a second time will take you back to the setting screen Pressing the On Off Button a third time will shut the device off QUICK TIP End of session At the end of yo...

Страница 8: ...xplosive Strength and Endurance training programs A The highest level possible but bearable in order to recruit the largest possible numbers of muscle fibers The contractions must be powerful without ever becoming uncomfortable Q What intensity should be reached for the special programs Potentiation and Active Recovery A These programs do not require the use of maximum current levels However the i...

Страница 9: ...ke you beyond what regular exercises can do 15 POTENTIATION PROGRAM Warm up your muscles quickly and prepare them to work faster and easier at the beginning of training or physical activity WARM UP BETTER ENDURANCE PROGRAM Delay muscle fatigue during activity Muscles perform more work with less oxygen Improves overall endurance RESISTANCE PROGRAM Increase strength of slow twitch muscles Improves p...

Страница 10: ... aerobic type of metabolism in muscle fibers and increases their capillary blood supply It trains the muscles to maintain effort for longer periods as it is required in endurance sports like marathons long distance bicycling etc Impulse sequences in this program are designed to improve the body s resistance to fatigue improved endurance The program targets very slow twitch muscle fibers Active Rec...

Страница 11: ...tioning Setting the appropriate intensity level See Intensity The Key to an Effective Workout on page 12 and the Contour Intensity Scale on page 13 19 Progression through the levels In general we don t recommend that you go through the different program levels quickly with the intention of getting to level 5 as soon as you can The levels represent progress in training and you should provide time f...

Страница 12: ...in Explosive Strength 5 120 Hz 3 sec 34 sec 4 Hz 20 min 35 min There is a ramp up of 2 seconds prior to the contraction time and a ramp down of one second after the contraction The average rest frequency is 4Hz Hz of stimulations per second e g 1 Hz 1 stimulation second 50 Hz 50 stimulations second 21 Phase Effect on Muscles Frequency Active Recovery Spasm Reduction Relaxation 1 3 Hz Endorphin Rel...

Страница 13: ...l pads won t stick to the belt even though they are relatively new A Make sure you remove the protective cover entirely before mounting the gel pad on the Contour Ab Belt Make sure that you are placing the silver side of gel pad electrodes to the BACK SIDE of the Ab Belt There will be outlines showing where to place the gel pad electrodes Once the gel pads have been applied to the Ab Belt leave th...

Страница 14: ...feel tingling in my legs during a session What can I do A The Contour electronic muscle stimulator emits an electrical frequency that will stimulate nerves motor neurons that are located beneath the gel pads Since the gel pads are larger it is possible to have some minor stimulation of other nerves located near the site of the gel pads Re adjust the gel pads slightly away from the area being unint...

Страница 15: ...l pad electrode is making good contact wIth your clean dry skin The battery symbol is flashing The batteries are low Replace all 4 AAA batteries Never mix old and new batteries The contractions are weak even though the intensity is high The batteries are low Replace all 4 AAA batteries Never mix old and new batteries The Gel Pads have lost conductivity Replace Gel Pads The pads are positioned impr...

Страница 16: ...he magnitude of electrical charge microcoulombs applied In the case of rectangular pulses the magnitude of charge is calculated by multiplying the duration of the pulse by the intensity of the current When stimulation is performed using a known pulse duration the intensity expressed in milliamperes or mA is a reproducible indicator of the degree of stimulation Definition of terms 29 Feature Contou...

Страница 17: ...50Hz 50Hz 50Hz DC Component Approx 32 µC 16 µC 11 µC Positive Pulse Width 300 µs 300 µs 300 µs 31 Care and cleaning To clean the Contour M2 device use a soft cloth lightly dampened with an alcohol based cleaning product that does not contain solvents Solvents could damage the plastic parts especially the display Your Contour M2 device is designed to be maintenance free There are no user serviceabl...

Страница 18: ...61000 4 6 3 V rms 150 kHz to 80 MHz outside ISM bands 3 V rms Radiated RF IEC 61000 4 3 3 V m 80 MHz to 2 5 GHz 3 V m 80 MHz to 800 MHz 800 MHz to 2 5 GHz 33 Where P is the maximum output power rating in the transmitter in watts W according to the transmitter manufacturer and d is the recommended separation distance in meters m Field strengths from fixed RF transmitters as determined by an electro...

Страница 19: ...s which may not operate properly when the electrical stimulation device is in use Operation in close proximity e g 1 yard to shortwave or microwave therapy EQUIPMENT may produce instability in the STIMULATOR output Applications with electrodes near the rib cage may produce heart ventricle fibrillation Application of electrodes near the thorax may increase the risk of cardiac fibrillation And injur...

Страница 20: ...or with OXYGEN OR NITROUS OXIDE Environmental conditions for transport and storage specified in accompanying documents and marked on packaging min temp and humidity Do not stand while you are stimulating the lower body as it may cause you to fall 37 POTENTIAL ADVERSE REACTIONS Skin irritation and burns beneath the electrodes have been reported with the use of powered muscle stimulators In order to...

Страница 21: ...ctrode set as skin reaction or cross contamination may occur Do not immerse the electrodes into water Do not apply a solvent of any kind onto the electrodes Always stop the Contour Technology Muscle Stimulator before removing or moving the electrodes Wash skin and dry skin thoroughly before applying the electrodes Apply the whole surface of the electrodes firmly to the skin Do not use electrodes t...

Отзывы: