background image

  Installation 

 

CPSN-SSI-4C Reference Manual (Hardware) 

  26 

 

3.

 

Cable Connection 

1. 

 

Thermocouple Input 

 

Thermocouple Input Cable 

Use the thermocouple input cable described below. 

 

Cable 

Thermocouple and compensating wire. 

Applicable wire 

AWG28 - 16 

Cable Length 

Vary according to the environment where the product is used. 

 

*Refer to 

"

Interface Connector

"

 in the 

page 18

 for details of the thermocouple input connector and 

pin assignment.   

 

 

Thermocouple Input 

Input type of thermocouple is differential input and four channels are provided for the product. 
Compatible thermocouples types are K, J, E, N, T, R, and S. 
Setting thermocouples type requires software command. (Default :K type) 
The measuring temperature range per thermocouples type is listed below. 
Even if the measuring temperature range is exceeded, it is possible to measure up to the measuring 
temperature limit, however the temperature tolerance may exceed the specified value. 

 

Measuring temperature range 

 

Thermocouples type 

Measuring temperature range 

-100°C - 1372°C 

-100°C - 1200°C 

-100°C - 1000°C 

-100°C - 1300°C 

-100°C - 400°C 

0°C - 1768°C 

0°C - 1768°C 

 

 

 

Содержание CPSN-SSI-4C

Страница 1: ...Manual Hardware CONTENTS Thermocouple Module CPSN SSI 4C Introduction 3 Safety Precautions 9 Product Nomenclature and Function 15 Installation 19 Appendix 29 Optional Products 34 Customer Support and...

Страница 2: ...Function 15 1 Nomenclature of Product Components 16 2 Description of Product Components 17 1 LED Indicator 17 2 Interface Connector 18 Installation 19 1 Install the Product 20 1 Installation Condition...

Страница 3: ...3 Introduction This section provides necessary information of the product such as the outline bundled items and manuals before actual use...

Страница 4: ...en operating the I O module with CPU Unit This describes the hardware aspects such as functions and settings of the I O module Download from the Contec website PDF CPU Unit Reference Manual Hardware R...

Страница 5: ...t visit our website to check the firmware version of the CPU unit and update to the latest one if necessary Updating firmware to the latest version will resolve troubles and stabilize the operation Do...

Страница 6: ...erence Manual Hardware 6 3 About the Product This product is an expansion I O module that adds a thermocouple interface to the CPU unit of the CONPROSYS nano series Four channels of differential input...

Страница 7: ...ion externally and temperature can be measured simply by connecting a thermocouple LED indicator for status check The error status of thermocouple such as disconnected not connected or sensor circuit...

Страница 8: ...The product consists of the items listed below Check with the following list that your package is complete If you discover damaged or missing items contact your retailer Product 1 10 pin Connector at...

Страница 9: ...9 Safety Precautions Understand the following definitions and precautions to use the product safely Never fail to read them before using the product...

Страница 10: ...derstand the meanings of these labels to operate the equipment safely DANGER DANGER indicates an imminently hazardous situation which if not avoided will result in death or serious injury WARNING WARN...

Страница 11: ...power medical equipment or other applications that require a very high level of reliability Do not use the product in such applications If using the product in applications where safety is critical su...

Страница 12: ...generation unplug the power cables of the CPU unit immediately and contact your retailer To clean the product wipe it gently with a soft cloth dampened with either water or mild detergent Do not use...

Страница 13: ...with industrial alcohol Use the product in accordance with the described installing method conditions and specifications If the product is used outside the specified range the protection performance...

Страница 14: ...inst harmful interference when the equipment is operated in a commercial environment This equipment generates uses and can radiate radio frequency energy and if not installed and used in accordance wi...

Страница 15: ...15 Product Nomenclature and Function This section describes product component names and their functions and pin assignment of each connector...

Страница 16: ...Nomenclature of Product Components Component names of the product are shown in the figure below No Name Function 1 LED Indicator This indicates status of the product 2 Interface Connector Connector f...

Страница 17: ...ribed below Color and Description LED Color Display Description 0 3 Green ON LED indicator to check the status It indicates the product operates properly Flash LED indicator to check the status It ind...

Страница 18: ...mocouple input terminal negative side of channel 0 3 TC1 Thermocouple input terminal positive side of channel 1 4 TC1 Thermocouple input terminal negative side of channel 1 5 AGND This is an analog gr...

Страница 19: ...19 Installation This section describes how to mount the product on a DIN rail or the wall and to connect to an external device with a cable...

Страница 20: ...shown below Other orientations could cause problems such as malfunction due to inadequate heat dissipation Vertical installation Installation angle 0 Vertical installation with an angle of 90 to the l...

Страница 21: ...tion angle for this product is 0 The temperature specifications of the product vary between the case when the installation angle is 90 to the left right and the case when a plane installation is used...

Страница 22: ...ction may occur if there is other device generating high heat the radiation will influence the product to increase its temperature Do not install this product into the fully sealed space except the ca...

Страница 23: ...R LED of the CPU unit is turned off before installing or removing the module Installation procedure 1 Paying attention to the horizontal orientation of the I O module insert the resin groove into the...

Страница 24: ...4C Reference Manual Hardware 24 Removal procedure 1 Press in the hooks at the top and bottom of the I O module and then pull it toward you and out of the CPU unit that has been installed on a DIN rail...

Страница 25: ...g the orange part in the center of the connector with a precision screwdriver or a similar tool insert the cable into the round connection hole After that release the pushed part to fix the cable in p...

Страница 26: ...ut type of thermocouple is differential input and four channels are provided for the product Compatible thermocouples types are K J E N T R and S Setting thermocouples type requires software command D...

Страница 27: ...ample of thermocouple connection The following figure shows an example of thermocouple connection Connect the positive and negative terminals of each thermocouple to the positive and negative sides of...

Страница 28: ...ation CPSN SSI 4C Reference Manual Hardware 28 Example of thermocouple connection Shielded CAUTION When using the product in an overly noisy environment use a shielded thermocouple and earth ground th...

Страница 29: ...29 Appendix This section lists the specifications and the physical dimensions of the product and the details of model name...

Страница 30: ...d temperature 0 12 Cold junction sensor Integrated Cold junction tolerance 2 3 Within 3 6 C Vertical installation Installation angle 0 Isolation Bus isolation Isolation withstand voltage 500VDC Connec...

Страница 31: ...ce Line noise Signal Line 1kV IEC61000 4 4 Level 3 EN61000 4 4 Level 3 Static electricity resistance Touch 4kV IEC61000 4 2 Level 2 EN61000 4 2 Level 2 Air 8kV IEC61000 4 2 Level 3 EN61000 4 2 Level 3...

Страница 32: ...Appendix CPSN SSI 4C Reference Manual Hardware 32 2 Physical Dimensions Physical dimensions of CPSN SSI 4C Physical dimensions of CPSN SSI 4C with connector attached...

Страница 33: ...PSN SSI 4C Reference Manual Hardware 33 3 The Details of Model Name Details of the model name are described below No Item Description 1 Interface SSI Sensor Input 2 Channel 4 4 Channels 3 Function C T...

Страница 34: ...34 Optional Products This section lists optional items that can be used along with the product...

Страница 35: ...1 Remote I O CPU unit CPSN MCB271 1 041 Remote I O CPU unit LAN 2 channel model CPSN PCB271 S1 041 CODESYS Modbus Master CPU unit DIN rail fitting power supply CPS PWD 30AW24 01 Fitting power supply 3...

Страница 36: ...36 Customer Support and Inquiry CONTEC provides the following support services for you to use CONTEC products more efficiently and comfortably...

Страница 37: ...he useful information including product manuals that can be downloaded through the CONTEC website Download https www contec com download You can download updated driver software firmware and different...

Страница 38: ...Revision History CPSN SSI 4C Reference Manual Hardware 38 Revision History MONTH YEAR Summary of Changes July 2019 The First Edition...

Страница 39: ...0025 Japan https www contec com No part of this document may be copied or reproduced in any form by any means without prior written consent of CONTEC CO LTD CPSN SSI 4C Reference Manual Hardware NA06...

Отзывы: