Installation
—
CPSN-SSI-4C Reference Manual (Hardware)
—
26
—
3.
Cable Connection
1.
Thermocouple Input
Thermocouple Input Cable
Use the thermocouple input cable described below.
Cable
Thermocouple and compensating wire.
Applicable wire
AWG28 - 16
Cable Length
Vary according to the environment where the product is used.
*Refer to
in the
page 18
for details of the thermocouple input connector and
pin assignment.
Thermocouple Input
Input type of thermocouple is differential input and four channels are provided for the product.
Compatible thermocouples types are K, J, E, N, T, R, and S.
Setting thermocouples type requires software command. (Default :K type)
The measuring temperature range per thermocouples type is listed below.
Even if the measuring temperature range is exceeded, it is possible to measure up to the measuring
temperature limit, however the temperature tolerance may exceed the specified value.
Measuring temperature range
Thermocouples type
Measuring temperature range
K
-100°C - 1372°C
J
-100°C - 1200°C
E
-100°C - 1000°C
N
-100°C - 1300°C
T
-100°C - 400°C
R
0°C - 1768°C
S