5.7    In boot-strap state,press button ,and the device is reset. 

  Fingernails and the luminescent tube should be on the same side.

 

6 Maintain, Transport and Storage 

6.1 Cleaning and disinfection 

The  device  must  be  turned  off  before  cleaning,  and  it  should  not  be  immersed  into 

liquid. 

Please take out the internal battery before cleaning, do not immerse it into liquid. 

Use 75% alcohol to wipe the device enclosure, and use liquid soap or isopropanol to 

wipe the watchband  for disinfection, nature dry or clean it with clean and soft cloth. 

Do not spray any liquid on the device directly, and avoid liquid penetrating into the 

device. 

6.2 Maintenance 

A. Check the main unit and all accessories periodically to make sure that there is no 
visible  damage  that  may  affect  patient’s  safety  and  monitoring  performance.  It  is 
recommended  that  the  device  should  be  inspected  weekly  at  least.  When  there  is 

obvious damage, stop using it. 

B. Please clean and disinfect the device before/after using it according to the User 

Manual (6.1). 

C. Please replace the batteries in time when low-battery appears. 

D.Please take out the batteries if the device is not used for a long time. 

E.

 

The device need not to be calibrated during maintenance. 

6.3 Transport and Storage 

A. The packed device can be transported by ordinary conveyance or according to 

transport contract. During transportation, avoid strong shock, vibration and splashing 

with rain or snow, and it can not be transported mixed with toxic, harmful, corrosive 

material. 

B.  The  packed  device  should  be  stored  in  room  with  no  corrosive  gases  and  good 

ventilation. Temperature: -40°C~+60°C; Relative humidity: ≤95%. 

7 Troubleshooting 

Trouble 

Possible Reason 

Solution 

The SpO

2

 

and 

Pulse Rate 

can not be 

displayed 

normally

 

1)

 

The finger is not properly 

inserted. 

2)

 

The finger is shaking or 

the patient is moving. 

3)

 

The device is not used in 

environment required by 

the manual. 

4)

 

The device works 

abnormally. 

1)

 

Please insert the finger 

properly and measure 

again. 

2)

 

Let the patient keep 

calm.   

3)

 

Please use the device 

in normal 

environment. 

4)

 

Please contact the 

after-sales. 

The SpO

2

 

and 

Pulse Rate 

are not 

displayed 

stably

 

1)

 

  The  finger  is  not  placed 

inside deep enough. 

2)

 

  The finger is shaking or the 

patient is moving. 

1)

 

  Place the finger 

properly and try again.   

2)

 

  Let the patient keep 

calm   

The device 

can not be 

turned on

 

1)

 

The battery is drained 

away or almost drained 

away. 

2)

 

The battery is installed 

incorrectly. 

3)

 

The device’s malfunction. 

1)

 

Please change 

batteries. 

2)

 

Please Install the 

battery again. 

3)

 

Please contact the 

local service center. 

The display is 

off suddenly

 

1)

 

The device enters into the 

energy saving mode. 

2)

 

Low battery. 

3)

 

The device works 

abnormally. 

1)

 

Normal. 

2)

 

Please charge the 

battery. 

3)

 

Please contact the 

after-sales. 

8 Key of Symbols 

Symbol 

Description 

 

Type BF 

 

Refer to instruction manual/booklet 

%SpO

2

 

The pulse oxygen saturation(%) 

PRbpm

 

Pulse rate (bpm) 

 

The battery voltage indication is deficient   

(change the battery in time avoiding the inexact measure) 

 

1. No finger inserted 

2. An indicator of signal inadequacy 

 

Battery positive electrode 

 

Battery cathode 

 

Exit standby mode 

SN 

Serial number         

 

Alarm inhibit 

 

WEEE (2002/96/EC) 

IP22

 

Ingress of liquids rank 

 

Storage and Transport Temperature limitation 

 

Storage and Transport Humidity limitation 

 

Storage and Transport Atmospheric pressure limitation 

 

This side up 

 

Fragile, handle with care 

 

Keep dry     

 

Recyclable 

 

European Representative 

 

This  item  is  compliant  with  Directive  93/42/EEC  of  14 

june  1993  concerning  medical  devices;  Including,  at  21 

march  2010,  the  amendments  by  Council  Directive 

2007/47/EEC. 

Note

Your device may not contain all the following symbols.

 

9 Function Specification 

SpO

2

 [see note 1] 

Display range 

0% ~ 99% 

Measured range 

0% ~ 100% 

Accuracy [see note 2] 

70%~100%: ±2%; 

0%~69%: unspecified. 

Resolution   

1% 

PR 

Display range 

30 bpm ~ 250 bpm 

Measured range 

30 bpm ~ 250 bpm 

Accuracy 

±2 bpm or ±2%, whichever is greater. 

Resolution   

1 bpm 

Accuracy  under  low 

perfusion [see note 3] 

Low perfusion 0.4%: 

SpO

2

: ±4%;   

PR: ±2 bpm or ±2%, whichever is greater 

Light interference 

Under normal and ambient light conditions, the SpO

2

 

deviation ≤ 1% 

Optical sensor [see note 4] 

Red light 

Wavelength: about 660 nm, optical output power: < 6.65 mW 

Infrared light 

Wavelength: about 905 nm, optical output power: < 6.75 mW 

Safety class 

Internally powered equipment, type BF applied part 

International 

Protection 

IP22 

Working voltage   

DC 2.6 V ~ 3.6 V 

Working current 

≤ 25 mA 

Power supply 

1.5V (AAA size) alkaline batteries × 2 or rechargeable battery 

Battery life 

Two batteries can work continually for 24 hours 

Dimension and Weight 

Dimension   

61 (L) × 36 (W) × 32 (H) mm 

Weight 

About 60g (with the batteries) 

 

Note 1: the claims of SpO

2

 accuracy shall be supported by clinical study measurements 

taken over the full range. By artificial inducing, get the stable oxygen level to the range 

of 70 % to 100 % SpO

2

, compare the SpO

2

 values collected by the  secondary standard 

pulse  oximeter  equipment  and  the  tested  equipment  at  the  same  time,  to  form  paired 

data, which are used for the accuracy analysis. 

There  are  12  healthy  volunteers  (male:  6.  female:  6;  age:  18~45;  skin  color:  black:  2,  light:  8, 

white: 2) data in the clinical report. 

Note  2:  because  pulse  oximeter  equipment  measurements  are  statistically  distributed, 

only  about  two-thirds  of  pulse  oximeter  equipment  measurements  can  be  expected  to 

fall within ±Arms of the value measured by a CO-OXIMETER.

 

Note  3:  percentage  modulation  of  infrared  signal  as  the  indication  of  pulsating  signal 

strength, patient simulator has been used to verify its accuracy under conditions of low 

perfusion. SpO

2

 and PR values are different due to low signal conditions, compare them 

with the known SpO

2

 and PR values of input signal. 

Note  4:  optical  sensors  as  the  light-emitting  components,  will  affect  other  medical 

devices applied the wavelength range. The information may be useful for the clinicians 

who  carry  out  the  optical  treatment.For  example,  photodynamic  therapy  operated  by 

clinician.

 

Note 5: Patient simulator has been used to verify the pulse rate accuracy, it is stated as 

the root-mean-square  difference  between  the  PR  measurement  value  and  the  value  set 

by simulator.

 

EMC 

Note:

 

 

The  device  is  subject  to  special  EMC precautions and it must  be  installed  and  used in 

accordance with these guidelines. 

 

The  electromagnetic  field  can affect  the device  performance,  so  other  equipment  used 

near  the  device  must  meet  the  corresponding  EMC  requirements.  Mobile  phones, 

X-rays or MRI devices are possible interference source, as they can emit high-intensity 

electromagnetic radiation. 

 

Refer  to  above  chapters  for  the  minimum  value  of  user’s  physiological  signal. 

Inaccurate result  will appear  when the  device  operates  with the  values  lower  than  the 

descriptions in above chapter 

 

The use of ACCESSORIES, transducers and cables other than those specified, with the 

exception  of  transducers  and  cables  sold  by  the  MANUFACTURER  of  the  ME 

EQUIPMENT  or  ME  SYSTEM  as  replacement  parts  for  internal  components,  may 

result in increased EMISSIONS or decreased IMMUNITY of the ME EQUIPMENT or 

ME SYSTEM. 

 

The device should not be used adjacent to or stacked with other equipment and that if 

adjacent or stacked use is necessary, it should be observed to verify normal operation in 

the configuration in which it will be used. 

 

Devices or systems may still be interfered by other equipment, even if other equipment 

meets the requirements of the corresponding national standard. 

 

Basic performance: SpO

2

 measured range: 70% ~ 100%, absolute error: ±2%; PR 

measured range: 30 bpm ~ 250 bpm, accuracy: ±2 bpm or ±2%, whichever is greater.   

Appendix 1

 

 

Guidance and manufacture's declaration 

Guidance and manufacture’s declaration – electromagnetic emissions- 

for all EQUIPMENT and SYSTEMS   

Guidance and manufacture’s declaration – electromagnetic emission 

The

 Pulse Oximeter

 is intended for use in the electromagnetic environment specified below. 

The  customer  of  the  user  of  the

  Pulse  Oximeter

  should  assure  that  it  is  used  in  such and 

environment. 

Emission test 

Compliance 

Electromagnetic environment – guidance 

RF emissions 

CISPR 11 

Group 1 

The

  Pulse  Oximeter 

uses  RF  energy  only  for  its 

internal  function.  Therefore,  its  RF  emissions  are 

very low and are not likely to cause any interference 

in nearby electronic equipment. 

RF emission 

CISPR 11 

Class B 

The

  Pulse  Oximeter 

is  suitable  for  use  in  all 

establishments,  including  domestic  establishments 

and  those  directly  connected  to  the  public 

low-voltage  power  supply  network  that  supplies 

buildings used for domestic purposes. 

 

Guidance and manufacture’s declaration – electromagnetic immunity – 

for all EQUIPMENT and SYSTEMS 

Guidance and manufacture’s declaration – electromagnetic immunity 

The

 Pulse Oximeter 

is intended for use in the electromagnetic environment specified below. 

The  customer  or  the  user  of

  Pulse  Oximeter 

should  assure  that  it  is  used  in  such  an 

environment. 

Immunity 

  test 

IEC 

60601 

  test level 

Complia

nce level 

Electromagnetic environment - 

guidance 

Electrostatic 

discharge (ESD) 

IEC 61000-4-2 

8 kV 

contact 

15 kV air 

8 kV 

contact 

15 kV air 

Floors should be  wood, concrete  or 

ceramic  tile.  If  floor  are  covered 

with  synthetic  material,  the  relative 

humidity should be at least 30%. the 

manufacturer  may  recommend  the 

ESD  precautionary  procedures  to 

user. 

Power frequency 

(50Hz) magnetic 

field 

IEC 61000-4-8 

30A/m 

30A/m 

Power frequency magnetic fields 

should be at levels characteristic of 

a typical location in a typical 

commercial or hospital 

environment. 

 

 

Guidance and manufacture’s declaration – electromagnetic immunity – 

for EQUIPMENT and SYSTEMS that are not LIFE-SUPPORTING 

Guidance and manufacture’s declaration – electromagnetic immunity 

The

 Pulse Oximeter

 

is intended for use in the electromagnetic environment specified below. 

The  customer  or  the  user  of

  Pulse  Oximeter 

should  assure  that  it  is  used  in  such  an 

environment.

 

Immunity 

test 

IEC 60601 

test level 

Compliance 

level 

Electromagnetic environment - guidance 

 

 

 

 

 

Conducted 

RF IEC 

61000-4-6 

 

 

 

 

 

 

Radiated 

RF IEC 

61000-4-3 

 

 

 

 

 

 

 

3V(0.15MH

z–80MHz),

6V(in ISM 

bands 

between 

0.15MHz 

and 

80MHz) 

 

10 V/m 

80  MHz  to 

2.7GH 

 

 

 

 

 

 

3V(0.15MHz
–80MHz),6V
(in ISM 

bands 

between 

0.15MHz and 

80MHz) 

 

 

10 V/m 

Portable  and  mobile  RF  communications 

equipment should be used no closer to any 

part of the Pulse Oximeter, including cables, 

than  the  recommended  separation  distance 

calculated  from  the  equation  applicable  to 

the frequency of the transmitter. 

Recommended separation distance 

 

 

 

80 MHz to 800 MHz 

 

                                800 MHz to 2.7 GHz 

 

Where  P  is  the  maximum  output  power 

rating  of  the  transmitter  in  watts  (W) 

according  to  the  transmitter  manufacturer 

and  d  is  the  recommended  separation 

distance in metres (m). 

Field  strengths  from  fixed  RF  transmitters, 

as  determined  by  an  electromagnetic  site 

survey,a should be less than the compliance 

level in each frequency range.b 

Interference  may  occur  in  the  vicinity  of 

equipment  marked  with  the  following 

symbol: 

NOTE 1  At 80 MHz and 800 MHz, the higher frequency range applies. 
NOTE 2  These guidelines may not apply in all situations. Electromagnetic propagation is 
affected by absorption and reflection from structures, objects and people. 

a    Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) 
telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV 
broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic 
environment due to fixed RF transmitters, an electromagnetic site survey should be 
considered. If the measured field strength in the location in which the Pulse Oximeter is 
used exceeds the applicable RF compliance level above, the Pulse Oximeter should be 
observed to verify normal operation. If abnormal performance is observed, additional 
measures may be necessary, such as reorienting or relocating the Pulse Oximeter. 
b    Over  the  frequency  range  150  kHz  to  80  MHz,  field  strengths  should  be  less  than  3 
V/m. 

 

Recommended separation distances between portable and mobile 

  RF communications equipment and the EQUIPMENT or SYSTEM –   

for EQUIPMENT or SYSTEM that are not LIFE-SUPPORTING 

Recommended separation distances between   

portable and mobile RF communications equipment and the

 Pulse Oximeter. 

 

The

  Pulse  Oximeter 

is  intended  for  use  in  an  electromagnetic  environment  in  which 

radiated RF disturbances are controlled. The customer or the user of the

 Pulse Oximeter 

can  help  prevent  electromagnetic  interference  by  maintaining  a  minimum  distance 
between portable and mobile RF communications equipment (transmitters) and the

 Pulse 

Oximeter 

as  recommended  below,  according  to  the  maximum  output  power  of  the 

communications equipment. 

 

Rated 

maximum 

output 

power of 

transmitter 

(W) 

Separation distance according to frequency of transmitter(m) 

150 kHz to 80 MHz 

P

V

d

1

5

.

3

 

80 MHz to 800 MHz 

P

E

d

1

5

.

3

 

800 MHz to 2.7 GHz 

P

E

d

1

7

 

0.01 

0.058 

0.035 

0.07 

0.1 

0.18 

0.11 

0.22 

0.58 

0.35 

0.7 

10 

1.83 

1.10 

2.21 

100 

5.8 

3.5 

For  transmitters  rated  at  a  maximum  output  power  not  listed  above,  the  recommended 
separation distance d in metres (m) can be estimated using the equation applicable to the 
frequency  of  the  transmitter,  where  P  is  the  maximum  output  power  rating  of  the 
transmitter in watts (W) according to the transmitter manufacturer. 
NOTE 1  At  80  MHz  and  800  MHz,  the  separation  distance  for  the  higher  frequency 
range applies. 
NOTE 2  These guidelines may not apply in all situations. Electromagnetic propagation 
is affected by absorption and reflection from structures, objects and people. 

 

P

V

d

1

5

.

3

P

E

d

1

5

.

3

P

E

d

1

7