4. Functions
AD12-64(PCI), AD12-16(PCI)
51
Start Condition
The condition for controlling the start of sampling can be selected from among software, input data
comparison and an external trigger. The conditions for controlling the start and stop of sampling are
completely independent of each other; they can be set separately.
-
Software
This product starts sampling and storing input data to memory immediately after the operation start
command is issued.
-
Input data comparison
When the operation start command is issued, this product compares the analog signal input through
a specified channel to the value of the preset comparison level. If the analog signal satisfies the
condition, this product starts storing input data.
Level comparison conditions are set as two conditions: level and direction.
The above sketch shows that the level comparison condition is satisfied in the rising direction.
The start condition is satisfied when the analog signal at the specified channel passes the
comparison level in the rising direction. Input data items are stored to memory, starting with those
at solid dots.
The above sketch shows that the level comparison condition is satisfied in the falling direction.
The start condition is satisfied when the analog signal at the specified channel passes the
comparison level in the falling direction. Input data items are stored to memory, starting with
those at solid dots.
If you set the level comparison directions to both directions, the start condition is satisfied when the
analog signal passes the level both in the rising and falling directions.
-
External trigger
This product starts waiting for an external control signal as soon as the operation start command is
output.
Sampling and data transfer to memory start when the external control signal (falling edge) is input.