Method programming

Method programming permits to store up to 9 different methods, each 

with 9 steps, of frequently used parameters in the non-volatile memory 

for future recall. Each step is able to recall a next one, providing a flex-

ible multiple step function for special techniques. The method mode also 

permits to program a linear voltage gradient for any step provided the 

limiting  current  or  power  is  not  attained.  Parameters  can  be  changed 

temporarily without interrupting the run.

1.   Verify if the instrument is switched off.

2.   Connect the electrophoresis tank to the instrument.

3.   Switch the instrument on. 

4.   Select the METHOD mode and press SET.

5.   Select the desired method and press SET.

6.   Select the desired step and press SET.

7.   Decide whether or not a voltage gradient has to be performed and 

press 

SET.

8.   Enter  the  desired  voltage  limit  and  press  SET  (=  the  voltage  end 

point in case of a gradient step).

9.   Enter the desired current limit and press SET (this limit is automati-

cally set to maximum and cannot be changed in case of a gradient 

step).

10. Enter the desired power limit and press SET (this limit is automati-

cally set to maximum and cannot be changed in case of a gradient 

step).

11. Enter the desired time-units (h of kVh) and press SET.

12. Enter the desired time and press SET.

13. Decide if the experiment should automatically proceed with a next 

step  (resuming  from  point  6)  or  stop  after  this  step.  Enter  the 

desired procedure and press 

SET.

14. Press RUN to start the experiment.

•  At any time, you can return in the stand-by mode by holding 

SET

 

pressed during 2 seconds.

•  At any time, you can avoid programming unnecessary parameters. 

To do so, simply press 

RUN 

instead of 

SET 

after entering the last 

relevant value and the instrument will immediately start the exper-

iment keeping the other parameters to their previous values.

•  To ensure that the desired parameter is kept constant, we suggest 

to pre-set all other parameters to maximum.

•  A blinking unit of the controlling  parameter appears when the regu-

lation is complete.

•  The timer will always count down.

•  If the timer is still running after switching on, the power supply will 

automatically proceed with the experiment for the remaining time. 

Press 

STOP 

if you want to go in stand-by during the run.

ENTER 

1

1

VOLTAGE: 

400V

----  STAND-BY  ----

METHOD

ENTER 

1

1

METHOD: 

1

ENTER 

1

1

STEP: 

1

ENTER 

1

1

GRADIENT: 

NO

ENTER 

1

1

CURRENT: 

300mA

ENTER 

1

1

POWER: 

50W

ENTER 

1

1

TIME UNIT: 

h

ENTER 

1

1

TIME: 

99:59h

ENTER 

1

1

END METHOD: 

YES

RUN 

1

99:59h

400V  300mA  50W

V

time

step 2

step 3

step 4

step 5

Содержание EV202

Страница 1: ...Consort MANUAL HANDLEIDING MODE D EMPLOI ANLEITUNG EV202 EV215 EV231 EV232 EV233 EV243 EV261 EV262 EV265 April 2006...

Страница 2: ...EV2xx www consort be 2 30...

Страница 3: ...IFICATE This instrument excluding all accessories is warranted against defective material and workmanship for a period of thirty six 36 months from the date of shipment ex fac tory Consort will repair...

Страница 4: ...ammation m thode 20 Programmation manuelle 21 Changer temporairement les param tres 21 Enregistrer des donn es 22 Alarmes 22 Inhaltsverzeichnis Hersteller 23 Einf hrung 23 Garantie 23 Service 23 Bedie...

Страница 5: ...for entering a value or for selecting a function SET ENTER Button for programming the desired parameters RUN STOP Button to start or stop an experiment Warranty This instrument excluding all accessor...

Страница 6: ...sis tank with safety cover Never make any other connections such as e g putting several power supplies in series or in parallel In order to prevent electric shock never open the back plate nor remove...

Страница 7: ...too low output current and press SET 8 Select POWER FAIL and press SET 9 Decide whether or not the instrument should proceed with the experiment for the remaining time after a power failure and press...

Страница 8: ...be changed in case of a gradient step 11 Enter the desired time units h of kVh and press SET 12 Enter the desired time and press SET 13 Decide if the experiment should automatically proceed with a ne...

Страница 9: ...e desired power limit and press SET 8 Enter the desired time units hours or kilovolthours and press SET 9 Enter the desired time and press SET pre set to if no timing is required 10 Press RUN to start...

Страница 10: ...d Questions at the end of this manual SHORT CIRCUIT Connected tank is short circuited or has a very low resistance check thoroughly the complete assembly Your power supply is NOT defective See Frequen...

Страница 11: ...ammeren RUN STOP Toets om het experiment te starten of te stoppen Fabrikant Consort nv Parklaan 36 Tel 32 14 41 12 79 B2300 Turnhout Fax 32 14 42 91 79 Belgi E mail info Consort be Waarborg Dit toeste...

Страница 12: ...oed ge soleerde elektroforesekamer met veiligheidsdeksel Maak nooit andere verbindingen zoals b v verschillende voedingen in serie of in parallel schakelen Om mogelijke elektrische schokken te vermijd...

Страница 13: ...druk op SET 8 Kies STROOMVERLIES en druk op SET 9 Beslis of het toestel al of niet automatisch moet verder gaan met het experiment voor de nog overblijvende tijd in geval van stroomverlies en druk op...

Страница 14: ...Kies de gewenste tijdseenheden h of kVh en druk op SET 12 Geef de gewenste tijd in en druk op SET 13 Beslis of het experiment automatisch met een volgende reeks para meters moet voortgaan ga terug na...

Страница 15: ...en druk op SET 7 Geef de gewenste vermogenslimiet in en druk op SET 8 Kies de gewenste tijdseenheden h of kVh en druk op SET 9 Geef de gewenste tijd in en druk op SET Stel in op wanneer geen tijdfunkt...

Страница 16: ...einde van deze hand leiding KORTSLUITING De aangesloten elektroforesekamer is kort gesloten of heeft een zeer kleine weerstand controleer grondig de gehele opzet De hoog spanningsbron is NIET defect Z...

Страница 17: ...tre d sir s RUN STOP Bouton pour d marrer ou arr ter l exp rience Fabricant Consort nv Parklaan 36 Tel 32 14 41 12 79 B2300 Turnhout Fax 32 14 42 91 79 Belgique E mail info Consort be Garantie Cet ap...

Страница 18: ...cordez les sorties uniquement une cuve d lectrophor se quip e d un couvercle de s curit Ne faites jamais d autres connexions comme p e mettre plusieurs g n rateurs en s rie ou en parall le Afin de pr...

Страница 19: ...t tre coup e en cas d un courant trop bas et appuyez sur SET 8 Choisissez PERTE DE COURANT et appuyez sur SET 9 D cidez si le g n rateur doit poursuivre automatiquement l exp rience pour le temps qui...

Страница 20: ...appuyez sur SET 12 Entrez le temps d sir et appuyez sur SET 13 D cidez si l exp rience doit se poursuivre automatiquement avec un pas suivant continuez partir du point 6 ou s arr ter ici Choisissez l...

Страница 21: ...consigne pour la puissance et appuyez sur SET 8 Choisissez les unit s de temps h ou kVh et appuyez sur SET 9 Entrez le temps d sir et appuyez sur SET pour liminer la tempo risation programmez un temps...

Страница 22: ...uestions la fin de ce mode d emploi COURT CIRCUIT La cuve branch e est court circuit e ou a une r sistance tr s basse v rifier profond ment l ensemble complet Votre g n rateur n est PAS d fectueux Voi...

Страница 23: ...grammierung der erw nschten Parameter RUN STOP Taste zum Starten oder Enden eines Experiments Hersteller Consort nv Parklaan 36 Tel 32 14 41 12 79 B2300 Turnhout Fax 32 14 42 91 79 Belgien E mail info...

Страница 24: ...dungen wie z B mehrere Netzger te in Serie oder parallel zu schalten sind unzul ssig Um einen Stromschlag zu vermeiden sollten Sie niemals die R ckplatte abneh men oder das Geh use ffnen Setzen Sie da...

Страница 25: ...en Ausgangsstrom und dr cken Sie SET 8 W hlen Sie STROMVERLUST und dr cken Sie SET 9 Entscheiden Sie sich ob das Experiment nach ein Stromverlust in der verbleibenden Zeit automatisch fortgesetzt werd...

Страница 26: ...12 W hlen Sie den gew nschten Zeitgrenzwert und dr cken Sie SET 13 Entscheiden Sie sich ob das Experiment automatisch mit einem n chsten Satz Parametern fortgesetzt werden soll wiederholen Sie Punkt 6...

Страница 27: ...gew nschten Leistungsgrenzwert und dr cken Sie SET 8 W hlen Sie die gew nschte Zeiteinheit h oder kVh und dr cken Sie SET 9 W hlen Sie den gew nschten Zeitgrenzwert und dr cken Sie SET Setzen Sie die...

Страница 28: ...ly Asked Questions am Ende dieser Anleitung KURZSCHLU Die angeschlossene Kammer ist kurzgeschlossen oder hat einen zu niedrigen Widerstand pr fen Sie das komplette System Das Netzger t ist NICHT defek...

Страница 29: ...d power then the maximum expected values Current Voltage Resistance Power Voltage x Current How to keep a constant current during a run Program the desired current and a higher voltage and power then...

Страница 30: ...9 set of parameters OUTPUTS 4 in parallel 4 mm sockets 4 in parallel 4 mm sockets 4 in parallel 4 mm sockets MIN LOAD RESISTANCE 5 15 70 NO LOAD DETECTION on off programmable on off programmable on of...

Отзывы: