15
3.3
Power Cord Requirements
100/120 Volt (Linvatec Cat. No. C7104)
Use only a listed (UL, CSA) detachable power cord manufactured to the following specifications.
•
Plug End
•
NEMA 5-15P hospital grade, 15 amps, 125V
•
Receptacle End
•
IEC 320/CEE-22, 6 amps, 250V/15 amps, 125V
•
Cord
•
UL style SJT, 14 AWG, 3 conductor
220/240 Volt (Linvatec Cat. No. C7105)
•
Plug End
•
Molded straight PVC plug with double grounding system
•
DIN 49441, CEE 7/U11, 10/16A, 250V
•
CEBEC, DEMKO, KEMA, NEMKO, OVE, SEMKO, VDE, UTE, FEMKO
•
Receptacle End
•
Molded straight PVC plug
•
DIN 49457, CEE 22/V, 10A, 250V
•
VDE, D, N, S, SEV, OVE, KEMA
•
Cord
•
PVC, 7.2mm diameter
•
10A, 250V
•
Conductors: 3 x 1 mm
2
•
Conductor Colors - brown, blue, green/yellow stripe
NOTE:
If the product is connected to a 240V system in the United States, then it must be
connected to a center tap system.
There are no toxic components contained in the Xenon Light Source. After the useful life of the
product, dispose of components and service parts properly.