14
2.
To operate the handpiece, move the safety/
direction lever to the “run” position and
depress the trigger.
CAUTION:
Do not operate the
Reciprocator Handpiece with the collet in the
open position or without a blade locked in the
collet. Damage will occur.
3.
To remove the blade:
(a) Ensure the safety/direction lever is in
the safe position.
(b) Rotate the blade locking knob
counterclockwise. The blade locking
collet will disengage. Remove the
blade.
3.0
MAINTENANCE
Regular and proper maintenance of your
PowerPro Handpieces are the best ways to pro-
tect your investment. It is essential that you have
your powered surgical instruments serviced as
scheduled in order to retain their optimum per-
formance and reliability, which will reward you
with safer, less problematic product perfor-
mance over time. The recommended mainte-
nance interval for service of your PowerPro
Handpieces is every 12 months.
Recommended care and handling of the instru-
ments include proper day-to-day operation,
cleaning, and sterilization which are extremely
important to ensure safe and efficient operation.
Your authorized Hall Surgical service depart-
ment is the most knowledgeable about the
PowerPro instruments and will provide compe-
tent and efficient service. Service at Linvatec/
Hall Surgical at the recommended service inter-
val is mandatory to keep your product warran-
ties in effect. Any services and/or repairs done
by any unauthorized repair facility may result in
reduced performance of the instruments or
instrument failure.