3.  Attach the relief valve to the outlet adapter on the tee-block. Remove 

the adapter if the ¼” female NPT thread port is required.

4.  Loosen the case of the relief valve approximately an 1/8 of a turn 

counter-clockwise by applying torque at the hex flats located at both 
ends of the valve.

5.  Insert the appropriate Allen wrench into the female NPT outlet port 

of the relief valve.

6.  Turn the Allen wrench clockwise to increase cracking pressure and 

counter clockwise to reduce.

7.  Tighten the case by reversing step four (4).
8.  Adjust the pressure of the 492-1312-01-580 to the desired relief point. 

Observe the point at which the relief exhausts.

9.  Open the needle valve a ¼ turn and back out the regulator-adjusting 

knob  until  the  outlet  gauge  pressure  reads  zero.  If  the  relief  valve 
exhausts  at  the  desired  pressure  go  to  step  eleven  (11).  If  further 
adjustment is required, proceed to step ten (10).

10.  Repeat  steps  four  (4)  through  nine  (9)  until  the  desired  cracking 

pressure is obtained.  

11.  Valve is ready for installation.

INSTALLATION

Note:  It  is  the  end  user's  responsibility  to  ensure  that  the  relief  system 
is  properly  designed  by  a  professional  engineer  to  meet  the  intended 
requirements.

1.  Be  sure  that  all  connections  are  secure  and  leak  tight. Teflon  tape 

should be used on pipe fittings, but avoid impinging on the gas stream. 
Before applying tape, inspect the NPT threads and if necessary, clean 
the fitting to remove any dirt or thread sealant that remains. Start the 
Teflon tape on the second thread as shown in Figure 3; make sure the 
tape does not overlap the end of the thread. As the tape is wrapped 
in the direction of the thread spiral (clockwise), pull tightly on the 
end of tape so that it conforms to the thread. Apply two overlapping 
layers and cut off excess tape.

Figure 3

Содержание 830-9412

Страница 1: ...1 800 225 0473 or 757 422 8330 Fax 757 422 3125 www concoa com 830 9412 830 9413 830 9414 Relief Valves December 2009 Supercedes Septemeber 2004 INSTALLATION AND OPERATION INSTRUCTIONS Certified ISO...

Страница 2: ...ther than CONCOA or a service facility designated by CONCOA CUSTOMER SERVICE In the event of equipment failure call CONCOA Customer Service Please be prepared to provide the model number and serial nu...

Страница 3: ...e Frequent activation will cause seat leak and require excessive maintenance DO NOT ADJUST WHILE UNDER PRESSURE 1 Maintainsystem operatingpressureatleast5 10 belowthesetpressure of the relief valve Op...

Страница 4: ...t the desired pressure go to step eleven 11 If further adjustment is required proceed to step ten 10 10 Repeat steps four 4 through nine 9 until the desired cracking pressure is obtained 11 Valve is r...

Страница 5: ...e outlet connection so as to prevent backpressure when relieving The piping termination point must be in a safe location and away from ignition sources The termination point must be positioned to prev...

Страница 6: ...THIS PAGE INTENTIONALLY LEFT BLANK...

Страница 7: ...other than Controls Corporation of America or if the equipment has been subject to abuse misuse negligence or accident Controls Corporation of America s sole and exclusive obligation and Buyer s sole...

Страница 8: ...Certified ISO 9001 ADI 9902 B Controls Corporation of America 1501 Harpers Road Virginia Beach VA 23454 To Order Call 1 800 225 0473 or 757 422 8330 Fax 757 422 3125 www concoa com...

Отзывы: