15
3.3 Wiring
The Dual O2/CO2 monitor requires a single, 3-wire shielded cable for analog output
and 24 VDC power input. A three-wire shielded cable; 3- conductor, 18 AWG stranded
General Cable E2203S.30.860, or equivalent is recommended for the connection. The
maximum permissible cable length is 0.62 miles (1 km). The analog out and VDC
power in connections are made on the terminal block inside the transmitter housing.
- Common
- Common
mA 4-20mA signal output to
PLC
+ 24VDC Power
+ 24VDC Power
Identification Legend
NOTE:
CONCOA has added additional contacts for +24VDC power and common to
accommodate additional wiring for remote horns and strobes
Pin #
Connection
Description
-
Common (Signal Ground)
0V
-
Common (Signal Ground)
0V
mA
Signal Out
DC 4-20mA Output
+
Power
DC + 24V Input
+
Power
DC + 24V Input