11
2: Specifications
NOTE:
For our continual product improvement, all specifications are subject to
change without notice.
2.1 Performance Specifications Oxygen
Sensor Type: Long Life Zirconium
Oxide Sensor Cell 0-25% Response Time:
Within 1 second of any change in O
2
.
Accuracy:
±1% of full scale
Fault Indicators:
Loss of VDC power (analog signal drops to 0 mA).
Sensor cell failure: Fault relay activated.
(Must have alarm relay option for cell
failure to operate)
Operating Temp:
-40° to 134°F (-40° to +55°C); consult CONCOA for lower or
higher operating temperatures.
Humidity:
0 to 95% RH; consult CONCOA for sensors which can operate in
100% condensing RH environments.
Environment:
Max. Altitude 2000 m, Pollution Degree 3, Intended for Indoor Use.
CE
EN 61000-3-2:2006 EMC, EN 61000-3-3:2008 EMC, EN61010-1-3-2013 LVD
2.2 Performance Specifications Carbon Dioxide sensor
Sensor Type:
Long Life NDIR Sensor 0 10,000ppm
Response Time:
Within 1 second of any change in CO
2
.
Accuracy:
±5% of measured value.
Fault Indicators:
Loss of VDC power (analog signal drops to 0 mA).
Sensor cell failure: Fault relay activated.
(Must have alarm relay option for cell failure to
operate)
Operating Temp:
32° to 122°F (0° to +50°C); consult CONCOA for lower or higher
operating temperatures.
Humidity:
0 to 95% RH.
Environment:
Max. Altitude 2000 m, Pollution Degree 3, Intended for Indoor Use.
CE
EN 61000-3-2:2006 EMC, EN 61000-3-3:2008 EMC, EN61010-1-3-2013 LVD
2.3 Gas Detection System
Transmitter: Microprocessor electronics with built-in 3-digit backlit LCD display Joystick
operated menus