Note that the voltage meter may indicate that some residual voltage is present on the main storage 
capacitor, even when the tester is first turned ON.  This is due to inherent charging of the internal 
capacitors.  Pressing the 

TRIGGER

 switch will discharge the capacitors (be sure not to touch the 

output and return leads when pressing the trigger switch). 

 

Note that the peak amplitude of the measured output waveform is proportional to the voltage that 
is read on the front panel of the MegaPulse, but it will always be somewhat lower.  This is because 
the meter on the MegaPulse is measuring the voltage on the main impulse storage capacitor.  This 
voltage will intentionally dissipate to some extent before reaching the output leads.  Therefore, it 
is important to measure the peak amplitude of the output waveform, and adjust the output of the 
MegaPulse accordingly. 

 

Determination  of  Passing  and  Failing  results  can  prove  difficult.    To  obtain  the  most  accurate 
results, it is generally necessary to perform multiple impulse tests on a few different test samples 
(that  have  adequate  insulation  to  pass  the  impulse  test).    Take  note  of  the  impulse  waveshape, 
amplitude, and duration.  Also note how much variance there is in the waveshape from test to test.  
Also  (if  possible),  perform  impulse  testing  on  some  test  samples  that  are  known  to  have 
inadequate  (or  damaged)  insulation.    Take  note  of  the  impulse  waveshape,  amplitude,  and 
duration, when an insulation breakdown occurs.  

 

 

Содержание MegaPulse Defib-5PF-002

Страница 1: ...MegaPulse IMPULSE TESTER Defib 5PF 002 Instruction Manual...

Страница 2: ...gn and ergonomic front panel and represents the latest in high voltage impulse testing To fully appreciate all the features of your new instrument we suggest that you take a few moments to review this...

Страница 3: ......

Страница 4: ...e Requirements 12 Fuse Replacement 12 Front and Rear Panel Features 13 Figure 5 MegaPulse Defib 5PF 002 Front Panel 13 Table 1 MegaPulse Defib 5PF 002 Front Panel 14 Figure 6 MegaPulse Defib 5PF 002 R...

Страница 5: ...5...

Страница 6: ...ept and creative at completing a path to ground Instructions should include a warning against any metal jewelry Operators should not allow others in the testing area especially when tests are being co...

Страница 7: ...ed Note that the CHARGE process will stop after 2 minutes if the TRIGGER button is not pressed The MegaPulse tester is designed to bleed the high voltage away after the test has concluded In order to...

Страница 8: ...ge capacitor This voltage will intentionally dissipate to some extent before reaching the output leads Therefore it is important to measure the peak amplitude of the output waveform and adjust the out...

Страница 9: ...eter Your tester is warranted for a period of one year upon shipment of the instrument to the original purchaser Specifications Specifications for the MegaPulse Defib 5PF 002 are listed in the table b...

Страница 10: ...10 Figure 1 Type A Diagram Figure 2 Type B Diagram...

Страница 11: ...11 Figure 3 Impulse waveform type A Figure 4 Impulse waveform type B...

Страница 12: ...r Remove the tester from its container and place it on a test bench AC Line Voltage Requirements AC line voltage requirements for your Tester are noted on the rear panel of the instrument Do not conne...

Страница 13: ...me familiar with the use of its controls indicators and connectors The front panel features of the MegaPulse are shown in Figure and described in Table 1 The rear panel features of the MegaPulse are s...

Страница 14: ...is lit for 2 minutes after the CHARGE button is pressed TRIGGER indicator will go out after pressing TRIGGER button TRIGGER and CHARGE Indicators will be blinking if the Interlock Switch is open Only...

Страница 15: ...se and required supply voltage for 115V input 5 Interlock Switch Emergency Stop Close Enables the tester buttons for operation Open Stops any process in the tester and disables the buttons The TRIGGER...

Страница 16: ...ment typically an oscilloscope with a high voltage probe Note that the RETURN lead is referenced to the chassis ground of the tester in this case we will connect channel 1 to the return of the probe c...

Страница 17: ...and connect the tester to a proper source of supply using the included 18AWG power supply cord 2 Plug the leads to the Outputs of the tester 3 Connect the ends of the test leads to the equipment under...

Страница 18: ...ce West USA is available Phone 800 748 6224 Hours 8 30 AM 4 30 PM Pacific Time Also available on our web site at www compwest com Contact Compliance West USA 650 Gateway Center Way Suite D San Diego C...

Страница 19: ...e corrected at no charge Mail the instrument post paid to the manufacturer Dated proof of purchase is required for all in warranty repairs The manufacturer is also available for calibration and or rep...

Страница 20: ...libration procedure must be performed by qualified personnel for more information contact Compliance West USA Voltage Stop Disable Keyboard Enable by Keyboard If the MegaPulse Defib 5PF 002 tester inc...

Страница 21: ...21 Appendix 1 Commands table...

Страница 22: ...22...

Страница 23: ...23...

Страница 24: ...24...

Отзывы: