Section 5

 

Maintenance and Calibration 

 

WARNING 

MAINTENANCE  AND  CALIBRATION  INSTRUCTIONS  ARE  FOR  QUALIFIED 
PERSONNEL  ONLY.    TO  AVOID  ELECTRIC  SHOCK,  DO  NOT  PERFORM  ANY 
SERVICING 

OTHER 

THAN 

THE 

CONTAINED 

IN 

THE 

OPERATING 

INSTRUCTIONS.

 

 

Introduction 

 

This  section  of  the  manual  contains  maintenance  information  for  the  MegaPulse  Defib-5PF-002 
impulse  tester.  A  1-year  calibration  cycle  is  recommended  to  maintain  the  specifications  of  the 
factory.    The  test  equipment  required  for  the  performance  test  is  a  digital  oscilloscope,  high 
voltage oscilloscope probe, digital meter and a high voltage probe. 

 

Service Information 

 

The MegaPulse tester is warranted to the original purchaser for a period of 1 year.  This warranty 
does not cover problems due to misuse or neglect.  Malfunctions which occur within the limits of 
the  warranty  will  be  corrected  at  no  charge.    Mail  the  instrument  post  paid  to  the  manufacturer.  
Dated proof of purchase is required for all in-warranty repairs.  The manufacturer is also available 
for  calibration  and  /  or  repair  of  instruments  that  are  beyond  their  warranty  period.    Contact  the 
manufacturer  for  a  cost  quotation.    Ship  the  instrument  and  your  remittance  according  to  the 
instructions given by the manufacturer.

 

 

General Maintenance 

 

 

To  avoid  contaminating  the  PWB  with  oil  from  your  fingers,  handle  it  by  the  edges  or  wear 
gloves.    If  the  PWB  becomes  contaminated,  refer  to  the  cleaning  procedures  given  later  in  this 
section

.

 

 

WARNING 

Dangerous  voltages  exist  when  energized.    Exercise  extreme  care  when  working  on  an 
energized circuit.

 

 

Cleaning 

 

Clean  the  front  panel  and  case with a mild solution of detergent and a damp sponge. Clean dust 
from the PWB with clean, dry, low pressure (<20 psi) 

 

CAUTION 

Do not use aromatic hydrocarbons or chlorinated solvents for cleaning.  These solutions will 

react with the plastic materials used in the instrument.

 

 
 
 
 

Содержание MegaPulse Defib-5PF-002

Страница 1: ...MegaPulse IMPULSE TESTER Defib 5PF 002 Instruction Manual...

Страница 2: ...gn and ergonomic front panel and represents the latest in high voltage impulse testing To fully appreciate all the features of your new instrument we suggest that you take a few moments to review this...

Страница 3: ......

Страница 4: ...e Requirements 12 Fuse Replacement 12 Front and Rear Panel Features 13 Figure 5 MegaPulse Defib 5PF 002 Front Panel 13 Table 1 MegaPulse Defib 5PF 002 Front Panel 14 Figure 6 MegaPulse Defib 5PF 002 R...

Страница 5: ...5...

Страница 6: ...ept and creative at completing a path to ground Instructions should include a warning against any metal jewelry Operators should not allow others in the testing area especially when tests are being co...

Страница 7: ...ed Note that the CHARGE process will stop after 2 minutes if the TRIGGER button is not pressed The MegaPulse tester is designed to bleed the high voltage away after the test has concluded In order to...

Страница 8: ...ge capacitor This voltage will intentionally dissipate to some extent before reaching the output leads Therefore it is important to measure the peak amplitude of the output waveform and adjust the out...

Страница 9: ...eter Your tester is warranted for a period of one year upon shipment of the instrument to the original purchaser Specifications Specifications for the MegaPulse Defib 5PF 002 are listed in the table b...

Страница 10: ...10 Figure 1 Type A Diagram Figure 2 Type B Diagram...

Страница 11: ...11 Figure 3 Impulse waveform type A Figure 4 Impulse waveform type B...

Страница 12: ...r Remove the tester from its container and place it on a test bench AC Line Voltage Requirements AC line voltage requirements for your Tester are noted on the rear panel of the instrument Do not conne...

Страница 13: ...me familiar with the use of its controls indicators and connectors The front panel features of the MegaPulse are shown in Figure and described in Table 1 The rear panel features of the MegaPulse are s...

Страница 14: ...is lit for 2 minutes after the CHARGE button is pressed TRIGGER indicator will go out after pressing TRIGGER button TRIGGER and CHARGE Indicators will be blinking if the Interlock Switch is open Only...

Страница 15: ...se and required supply voltage for 115V input 5 Interlock Switch Emergency Stop Close Enables the tester buttons for operation Open Stops any process in the tester and disables the buttons The TRIGGER...

Страница 16: ...ment typically an oscilloscope with a high voltage probe Note that the RETURN lead is referenced to the chassis ground of the tester in this case we will connect channel 1 to the return of the probe c...

Страница 17: ...and connect the tester to a proper source of supply using the included 18AWG power supply cord 2 Plug the leads to the Outputs of the tester 3 Connect the ends of the test leads to the equipment under...

Страница 18: ...ce West USA is available Phone 800 748 6224 Hours 8 30 AM 4 30 PM Pacific Time Also available on our web site at www compwest com Contact Compliance West USA 650 Gateway Center Way Suite D San Diego C...

Страница 19: ...e corrected at no charge Mail the instrument post paid to the manufacturer Dated proof of purchase is required for all in warranty repairs The manufacturer is also available for calibration and or rep...

Страница 20: ...libration procedure must be performed by qualified personnel for more information contact Compliance West USA Voltage Stop Disable Keyboard Enable by Keyboard If the MegaPulse Defib 5PF 002 tester inc...

Страница 21: ...21 Appendix 1 Commands table...

Страница 22: ...22...

Страница 23: ...23...

Страница 24: ...24...

Отзывы: