9
ITEM
NAME
FUNCTION
1
RESET
Button / Red Indicator
When lit, indicates that the HT-2800 is unarmed. This button must be pushed before the
TEST
Button is
functional. When the
RESET
Button is pressed, the red lamp goes out and the yellow
TEST
lamp is lit.
PRESSING THE
RESET
BUTTON AT ANY TIME STOPS TEST.
2
TEST
Button / Yellow
Indicator
When lit, indicates that the HT-2800 is ready to test the connected equipment. The yellow
TEST
indicator goes
out when the
TEST
Button is pressed.
3
Ground OK
/
Ground Open
LED's
Indicates result of continuity test between grounding pin of the line cord and exposed metal parts of the
equipment under test. If the ground circuit has more than three ohm resistance, or if the leads are not properly
connected, the red continuity LED will light, the internal buzzer will sound, and the test will be terminated. If the
ground circuit resistance is acceptable, the green continuity LED will light and the test will continue. If the
continuity test is defeated by the operator, both LED's will be lit for the duration of the test.
4
Excess Leakage
LED
Indicates failure of leakage current test. If leakage current between the primary circuit and ground is higher than
the preset value, the red LED will light, the internal buzzer will sound, and the test will be terminated.
5
Full Voltage
LED
If full voltage is successfully reached, the
Full Voltage
LED will light and the high voltage duration time starts.
6
Hipot Fail
LED
Indicates failure of high voltage test. If arcing or a flashover of the insulation system between primary parts and
ground is encountered, the red breakdown LED will light, the internal buzzer will sound, and the test will be
terminated.
7
Hipot Pass
LED
At the end of the preset high voltage duration time, if no insulation breakdowns are encountered, the green light
will light and the test will terminate.
8
Voltage Meter
Provides visual indication to the operator of the actual output voltage of the Tester. Used to set the automatic
voltage level during the setup procedure.
9
Ramp Bar Graph LED’s
Individual LED’s light in sequence from left to right as the test voltage ramps from zero to full voltage. LED’s
turn off (ramp back down) at end of test.
10
Return Lead Receptacle
Grounded banana plug receptacle. The 18 AWG Test Return Lead provided is connected here.
11
High Voltage Device Receptacle
NEMA 5-15R receptacle. For connection of the equipment under test.
Table 2. Controls, Indicators, Connectors - Model HT-2800 Front Panel