6
ELECTRICAL
Output
0-2800 Volts DC into NEMA 5-15R Receptacle.
Pigtail and Receptacle adapters available for most
configurations. Consult factory.
Leakage Current
0-5 mA
Pass/Fail Criteria:
Continuity:
3 ohm ± 5% * (defeatable)
Leakage Current:
User adjustable 1-5 mA
Dielectric Breakdown:
Separate high frequency detection circuit for
breakdown spike detection
Test Time:
User adjustable 1-60 sec.
Voltage Ramp Time:
User adjustable <2-10 sec.
Pass/Fail Repeatability
?
3%
Meter accuracy
+ 20V (0.75% of full-scale reading)
Duty cycle
100 %
Discharge Time
Voltage dependent; discharge progress displayed on
front panel meter
Test adjustments
Tamperproof operation available
ENVIRONMENTAL
Operating Temperature
15-40
?
C
Relative Humidity Range
0-90% non-condensing
GENERAL
Input power requirements
Model HT-2800: 114-127 volts, 50/60 Hz
Model HT-2800L: 98-120 volts, 50/60 Hz
Weight
10½ lbs
Dimensions
9¼ in. wide, 4¾ in. high, 10 in. deep.
SAFETY AGENCY TOPICS
Transformer Output
< 500VA
Visual Indication of Voltage Output Provided by front panel meter, directly connected to high
voltage output
Failure Indication
Audible, provided by internal buzzer
Visual, provided by red LED's on front panel
Test automatically terminated on failure
Continuity Test
Provided; 3 ohm pass/fail point*. May be defeated
for testing of products with two wire power supply
cords.
Leakage Test
Provided; 2 mA factory set pass/fail point, user
adjustable. 120kohm calibration resistor provided on
rear panel.
Operation Instructions
Provided on rear panel.
Table 1. HT-2800 Specifications
*Continuity Level can be adjusted from .5 to 5 ohms during the calibration procedure.