19
EN
7. EMC TABLE
The Compex stimulator needs special precautions regarding EMC and needs to be installed and put into
service according to the EMC information provided.
Portable and mobile RF communications equipment can affect the Compex stimulator.
The use of Accessories, transducers, and cables other than those specified by the manufacturer, may result
in increased Emissions or decreased Immunity of the Compex stimulator.
The Compex stimulator should not be used adjacent to or stacked with other equipment and if adjacent
or stacked use is necessary, the Compex stimulator should be observed to verify normal operation in the
configuration in which it will be used.
GUIDANCE AND MANUFACTURER’S DECLARATION – ELECTROMAGNETIC EMISSIONS
The Compex stimulator is intended for use in the electromagnetic environment specified below. The customer or the user of
the Compex stimulator should assure that it is used in such an environment.
EMISSION TESTS
COMPLIANCE
ELECTROMAGNETIC ENVIRONMENT
– GUIDANCE
Radiated Emissions CISPR 11
Group 1
The Compex uses RF energy only for
its internal function. Therefore, its RF
emissions are very low and are not
likely to cause any interference in
nearby electronic equipment.
Conducted Emissions CISPR 11
Class B
Harmonic emissions IEC 61000-3-2
Class A
The Compex is suitable for use in all
establishments, including domestic
establishments and those directly
connected to the public low-voltage
power supply network that supplies
buildings used for domestic purposes.
Voltage fluctuations IEC 61000-3-3
Complies