Connecting the unit electrically
Requirements
✓
Coolant hoses connected, see "Connecting the coolant hoses".
✓
Cooling water hoses connected, see "Connecting the cooling water hoses".
✓
Coolant added, see "Adding coolant".
✓
Unit cover removed, see "Removing and mounting the unit cover".
Procedure
•
Connect the unit to mains by inserting the mains plug or making a mains connection as
required by the particular periphery.
The unit is now connected electrically.
Figure 6: Mounting position of the terminal box
(1) Terminal box
6.4. Putting the unit into operation
NOTE
Danger of malfunction caused by faulty connections during initial operation!
Before switching on the unit make sure that:
All safety equipment of the unit is implemented and functional.
All connections were properly made.
Nobody is endangered by the start-up of the unit.
NOTE
Lack of coolant may impair cooling capacity and destroy the pump!
Only operate the unit with correct coolant level.
Check the coolant level regularly.