IE-SNC-T33(4)11+T73(4)10-07 

 

  3 

temperature unit:

 

°C 

pressure unit:

 

hPa 

preset computed value: 

dew point temperature 

correction for altitude: 

0 hPa

 (absolute pressure) 

 

Device installation 

Device is designed for wall mounting. There are two mounting holes at the sides of the case. It 

is NOT recommended to use the device for long time under condensation conditions. It could be the 
cause  of  water  steam  condensation  inside  the  sensor’s  cover  into  water  phase.  This  liquid  phase 
stays  inside  sensor’s  cover  and  can’t  escape  from  the  cover  easily.  It  can  dramatically  increase 
response time to relative humidity change. If water condensation occurs for longer time it can cause 
sensor damage. Similar effect can occur under water aerosol conditions. Don’t connect device while 
power supply voltage is on. Interconnection terminals for T3311, T3313, T7310, T3411 and T7410 
devices are accessible after unscrewing four screws and removing the lid. Lace the cable through a 
gland  at  the  case  wall.  Connect  the  cable  to  terminals  with  respecting  the  signal  polarity  (see 
figure). Terminals are self-clamping and can be opened by a suitable screwdriver. For the opening, 
insert the screwdriver to smaller terminal hole and lever by him. Do not forget to tighten glands and 
case lid with  inserted packing  after  cables  connecting.  It  is  necessary  for warranting of protection 
IP65.  Connect  complementary  female  connector  for  T3311L,  T3313L,  T7310L,  T3411L  and 
T7410L transmitters in accordance with the table in Appendix B of this manual. Working position 
is with the sensor filter downwards – see figure. 

Devices  T3311,  T3313  and  T7310  are  supplied  with connection  cable  equipped  with 

connector for connection to RS232 interface. For devices with RS485 output there is recommended 
to use shielded twisted copper cable, maximal length 1200m. The cable must be located at indoor 
rooms.  Nominal  cable  impedance  should  be  100 Ω,  loop  resistance  max.  240 Ω,  cable  capacity 
max. 65 pF/m. The cable should be led in one line, i.e. NOT to „tree“ or „star“. Termination resistor 
should be located at the end. For short distance other topology is allowed. Terminate the network by 
a  termination  resistor.  The  value  of  the  resistor  is  recommended  about  120 

.  For  short  distance 

termination resistor can be left out. Outside diameter of the cable for  T3411 and T7410 connection 
must be from 3 to 6.5 mm. Suitable cable is e.g. SYKFY 2x2x0.5 mm

2

, where one wire pair serves 

for  device  powering  and  the  other  pair  for  communication  link.  For  devices  T3311L,  T3313L, 
T7310L,  T3411L  and  T7410L  use  cable  with  respect  to  female  connector  parameters.  Do  NOT 
connect  shielding  at  connector  side.  The  cable  should  not  be  led  in  parallel  along  power  cabling. 
Safety distance is up to 0.5 m, otherwise undesirable induction of interference signals can appear. 

Electrical  system  (wiring)  may  do  only  worker  with  required  qualification  by  rules  in 

operation. 

 
 

Info mode 

If in doubt of setting of installed device, verification of its address is enabled even without 

using computer. Power should be connected. Devices with RS232 interface have address always set 
to one. 

Unscrew device cover and shortly press button next to connection terminals (jumper should 

be opened). Actual adjusted address of the device is displayed on LCD display at decimal base, for 
HWg-Poseidon’s  communication  protocol  there  is  shown  number  corresponding  with  ASCII 
address code. Next press of button exits info mode and actual measured values are displayed.  
 
Note:  No  measurement  and  communication  is  possible  during  info  mode.  If  device  stays  in info 
mode for longer than 15 s, device automatically returns to measuring cycle. 

Содержание T3311

Страница 1: ...lative humidity and other derived humidity values T3311 T3313 T3411 Programmable transmitter of atmospheric pressure temperature relative humidity and other derived humidity values T7310 T7410 with RS...

Страница 2: ...shown on dual line LCD display If there are two values displayed at one LCD line they are periodically switched between both readings with period of 4 seconds Display can be switched OFF totally too...

Страница 3: ...ices with RS485 output there is recommended to use shielded twisted copper cable maximal length 1200m The cable must be located at indoor rooms Nominal cable impedance should be 100 loop resistance ma...

Страница 4: ...4 IE SNC T33 4 11 T73 4 10 07 Dimensions T3311 T7310 Transmitter with RS232 Dimensions T3313 Transmitter with RS232...

Страница 5: ...IE SNC T33 4 11 T73 4 10 07 5 Dimensions T3411 T7410 Transmitter with RS485 Typical application wiring connection of terminals Power from communication port Power from external voltage source...

Страница 6: ...6 IE SNC T33 4 11 T73 4 10 07 Dimensions T3311L T3411L T7310L T7410L Connection see Appendix B Dimensions T3313L Connection see Appendix B...

Страница 7: ...ght bit data word in format 1 start bit data word 8 bit LSB first 2 stop bits1 without parity Device supports communication speed from 110Bd to 115200Bd Sent request and response have syntax ADDRESS O...

Страница 8: ...inH2O Int 10 kPa Int 100 Since FW version 02 44 there is possibility to read the all computed values from registers Variable Unit Address hex X Address dec X Format Size Status Dew point temperature...

Страница 9: ...ess and check sum is reset in accordance with values stored in the device Communication speed and check sum are possible to change only if jumper is closed Jumper closed and button pressed for longer...

Страница 10: ...g procedure to device s memory occurred or if damage of calibration data appeared At this state device does not measure and calculate values It is a serious error contact distributor of the device to...

Страница 11: ...state of value If selected pressure unit is mBar or oz in2 or mmHg or inH2O or kPa there is shown only value number without corresponding pressure unit C F DP reading next to this symbol is calculate...

Страница 12: ...Pa Range 600 8 70 450 0 17 72 240 9 139 2 60 00 1100 15 95 825 1 32 48 441 6 255 3 110 00 accuracy T 23 C 1 3 0 02 1 0 0 04 0 5 0 3 0 13 0 T 40 C 1 5 0 02 1 1 0 04 0 6 0 3 0 15 Else 2 0 0 03 1 5 0 06...

Страница 13: ...ve humidity including their accuracy you can exactly determine by the program Conversions It is free to download at http www cometsystem cz dnload conversions exe Response time air flow approximately...

Страница 14: ...ation It is not allowed to operate the device under conditions other than specified in technical parameters Devices are not designed for locations with chemically aggressive environment Temperature an...

Страница 15: ...cted to the bus by connecting the corresponding pins of CANON connector for more information see the operation manual for ELO E214 The ELO E06D converter is an optional accessory for connection of tra...

Страница 16: ...16 IE SNC T33 4 11 T73 4 10 07 Appendix B Connection of transmitters version TxxxxL...

Отзывы: