Dew point temperature
Range:
-60 to +80 °C (-76 to 176 °F)
Accuracy:
±1.5 °C (±2,7 °F) at ambient temperature T < 25°C (77°F) and relative
humidity RH >30%, for more details see graphs below
Specific humidity
2
Accuracy:
±2g/kg at air temperature T < 35°C (95°F)
Range:
0 to 550 g/kg
Mixing ratio
Chyba! Záložka není definována.
Accuracy:
±2g/kg at air temperature T < 35°C (95°F)
Range:
0 to 995 g/kg
Specific enthalpy
Chyba! Záložka není definována.
Accuracy:
± 3kJ/kg at air temperature T < 25°C (77°F)
Range:
0 to 995 kJ/kg
3
The values computed from ambient temperature and relative humidity including their
accuracy you can exactly determine by the program
Conversions
. It is free to
http://www.cometsystem.cz/dnload/conversions.exe
Response time
(air flow approximately 1 m/s):
temperature (temperature step 20°C):
- with stainless steel mesh sensor cover (F5200) t90 < 7 min
- with bronze sensor cover (F0000 - selectable option) t90 < 9 min
relative humidity: t90 < 30 s (humidity step 65 %RH, constant temperature)
atmospheric pressure
4
: t90 <44s
2
This value depends on the barometric pressure. If device doesn’t support pressure measurement, then
constant value stored inside device memory is used. Default value preset by manufacturer is 1013hPa and
can be changed by user’s software.
3
This maximum is reached under conditions about 70°C/100%RH or 80°C/70%RH