Comet T0310 Скачать руководство пользователя страница 9

IE-SNC-T03(4)10+T5410-06 

 

  9 

pressure  at  kPa (depended by device type).  For  atmospheric pressure correction to  altitude setting 
there is Users software TSensor. 

Jumper and button 

If  communication  with  HWg  Poseidon  unit  is  selected,  the  function  of  jumper  and  button  is  as 
follows: 

 

Jumper  opened  and  button  shortly  pressed  –  device  goes  to  Info  mode,  see  chapter  „Info 
mode“

 

Jumper  closed  and  button  pressed  for  longer  than  six  seconds  –  causes  restoring  of 
manufacturer  setting 

of  communication  protocol

,  i.e.  sets  Modbus  RTU  communication 

protocol,  device  address  sets  to  01h  and  communication  speed  to  9600Bd  –  after  button 
press  there  is  “dEF”  message  blinking  at  LCD  display.  Six  seconds  later  message  “dEF” 
stays shown, it means manufacturer setting 

of communication protocol

 is done. 

 

Error States of the device 

Device  continuously  checks  its  state  during  operation.  In  case  error  is  found  LCD  displays 
corresponding error code: 

Error 0 

First line displays „Err0“. 
Check  sum  error  of  stored  setting  inside  device’s  memory.  This  error  appears  if  incorrect  writing 
procedure to device’s memory occurred or if damage of calibration data appeared. At this state device 
does not measure and calculate values. It is a serious error, contact distributor of the device to fix. 

Error 1 

Measured  temperature  (no  pressure)  is  over  upper  limit  of  allowed  full  scale  range.  There  is 
a reading „Err1“ on LCD display. Value read from the device is +999.9 (for pressure reading there 
is +999.9 hPa correct value). 
This  state  occurs  if  measured  temperature  is  higher  than  approximately  600°C  (i.e.,  high 
 non-measurable resistance of temperature sensor, probably opened circuit). 

Error 2 

There is a reading „Err2“ on LCD display. Measured temperature is below lower limit of allowed 
full scale range or pressure measurement error occurred. Value read from the device is -999.9. 
This state appears in case of: 

 

Measured  temperature  is  lower  than  approximately  -210°C  (i.e.  low  resistance  of 
temperature sensor, probably short circuit). 

 

Measured  pressure  value 

with  added  correction  for  altitude

  is  outside  of  range  from 

300 hPa  to  1350 hPa  (from  4.351  PSI  to  19.580  PSI)  or  the  pressure  sensor  is  damaged. 
Please check setting of pressure correction for altitude with User’s software. 

Error 3 

There is a reading „Err3“ on LCD display. 
Error  of  internal  A/D  converter  appeared  (converter  does  not  respond,  probably  damage  of  A/D 
converter).  This  error  does  NOT  affect  pressure  measurement.  Measured  temperature  is  NOT 
measured. It is a serious error, contact distributor of the device. 

Error 4 

There  is  a  reading  „Err4“  on  LCD  display.  It  is  internal  device  error  during  pressure  sensor 
initialization.  Under  this  condition  device  does  NOT  measure  atmospheric  pressure.  Value  read 
from device is -999.9. Pressure sensor is probably damaged. It is a serious error, contact distributor 
of the device. 
 

Содержание T0310

Страница 1: ...nd improvement of their products Manufacturer reserves the right to make technical changes to the device without previous notice Misprints reserved Manufacturer is not responsible for damages caused b...

Страница 2: ...ware TSensor for setting of all device parameters recommended It is free to download at www cometsystem com It supports make the adjustment of the device too This procedure is described at file Calibr...

Страница 3: ...nection to RS232 interface For devices with RS485 output there is recommended to use shielded twisted copper cable maximal length 1200m The cable must be located at indoor rooms Nominal cable impedanc...

Страница 4: ...4 IE SNC T03 4 10 T5410 06 Dimensions T0410 T5410 Transmitter with RS485 Dimensions T0310L T0410L T5410L Connection see Appendix B...

Страница 5: ...IE SNC T03 4 10 T5410 06 5 Typical application wiring connection of terminals T0310 power from communication port T0310 power from external voltage source T0410 T5410 RS485...

Страница 6: ...tocols including examples of communication is available in individual document Description of communication protocols of Txxxx series which is free to download at www cometsystem com Note After switch...

Страница 7: ...R Serial number of device Lo 0x1036 4150 BCD BIN16 R Version of Firmware Hi 0x3001 12289 BCD BIN16 R Version of Firmware Lo 0x3002 12290 BCD BIN16 R Addition for device T5410 with atmospheric pressure...

Страница 8: ...pressed for longer than six seconds causes restoring of manufacturer setting of communication protocol i e sets Modbus RTU communication protocol device address sets to 01h and communication speed to...

Страница 9: ...it of allowed full scale range There is a reading Err1 on LCD display Value read from the device is 999 9 for pressure reading there is 999 9 hPa correct value This state occurs if measured temperatur...

Страница 10: ...ameters Ambient temperature RTD sensor Pt1000 3850ppm Measuring range 30 to 80 C Resolution of display 0 1 C Accuracy 0 4 C Atmospheric pressure T5410 only Unit hPa mBar PSI mmHg inHg inH2O oz in2 kPa...

Страница 11: ...ions other than specified in technical parameters Devices are not designed for locations with chemically aggressive environment Temperature sensor must not be exposed to direct contact with water or o...

Страница 12: ...the corresponding pins of CANON connector for more information see the operation manual for ELO E214 The ELO E06D converter is an optional accessory for connection of transmitter with RS485 interface...

Отзывы: