Comet T0310 Скачать руководство пользователя страница 4

 

IE-SNC-T03(4)10+T5410-06 

Dimensions 

– T0410, T5410 

 
 
 
 

Transmitter with RS485  

 

 

 
 

 
 
 

 
 

 
 
 
 
 
 
 
 

Dimensions 

– T0310L, T0410L, T5410L 

 

 

 
 

Connection:

 see Appendix B 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

 

 

 

Содержание T0310

Страница 1: ...nd improvement of their products Manufacturer reserves the right to make technical changes to the device without previous notice Misprints reserved Manufacturer is not responsible for damages caused b...

Страница 2: ...ware TSensor for setting of all device parameters recommended It is free to download at www cometsystem com It supports make the adjustment of the device too This procedure is described at file Calibr...

Страница 3: ...nection to RS232 interface For devices with RS485 output there is recommended to use shielded twisted copper cable maximal length 1200m The cable must be located at indoor rooms Nominal cable impedanc...

Страница 4: ...4 IE SNC T03 4 10 T5410 06 Dimensions T0410 T5410 Transmitter with RS485 Dimensions T0310L T0410L T5410L Connection see Appendix B...

Страница 5: ...IE SNC T03 4 10 T5410 06 5 Typical application wiring connection of terminals T0310 power from communication port T0310 power from external voltage source T0410 T5410 RS485...

Страница 6: ...tocols including examples of communication is available in individual document Description of communication protocols of Txxxx series which is free to download at www cometsystem com Note After switch...

Страница 7: ...R Serial number of device Lo 0x1036 4150 BCD BIN16 R Version of Firmware Hi 0x3001 12289 BCD BIN16 R Version of Firmware Lo 0x3002 12290 BCD BIN16 R Addition for device T5410 with atmospheric pressure...

Страница 8: ...pressed for longer than six seconds causes restoring of manufacturer setting of communication protocol i e sets Modbus RTU communication protocol device address sets to 01h and communication speed to...

Страница 9: ...it of allowed full scale range There is a reading Err1 on LCD display Value read from the device is 999 9 for pressure reading there is 999 9 hPa correct value This state occurs if measured temperatur...

Страница 10: ...ameters Ambient temperature RTD sensor Pt1000 3850ppm Measuring range 30 to 80 C Resolution of display 0 1 C Accuracy 0 4 C Atmospheric pressure T5410 only Unit hPa mBar PSI mmHg inHg inH2O oz in2 kPa...

Страница 11: ...ions other than specified in technical parameters Devices are not designed for locations with chemically aggressive environment Temperature sensor must not be exposed to direct contact with water or o...

Страница 12: ...the corresponding pins of CANON connector for more information see the operation manual for ELO E214 The ELO E06D converter is an optional accessory for connection of transmitter with RS485 interface...

Отзывы: