Protocol compatible with Advantech-ADAM standard  

Control units communicate on master-slave principle in half-duplex operation. Only master can 

send requests and only addressed device responds. During sending request any of slave devices should 
respond. During communication data is transferred in ASCII format (in characters). Each Byte is sent as 
two ASCII characters. Device supports communication speed from 1200Bd to 115200Bd, parameters of 
communication link are 1 start bit + eight bit data word (LSB first) + 1 stop bit, without parity. 

Jumper 

Jumper is located next to connection terminals. If communication protocol compatible with 

standard Advantech-ADAM is selected, its function is the following: 

 

If jumper during switching ON the power is closed, device always communicates with 
following parameters regardless stored setting in the device: 
communication speed 9600 Bd, without check sum, device address 00h 

 

If jumper during switching ON the power is not closed, device communicates in accordance 
with stored setting. 

 

If jumper is closed during device operation, device temporarily changes its address to 00h, 
it will  communicate  in  the  same  communication  speed  as  before  closing  jumper  and  will 
communicate without check sum. After jumper is opened setting of address and check sum 
is reset in accordance with values stored in the device. 

 

Communication speed and check sum are possible to change only if jumper is closed. 

 

Jumper  closed  and  button  pressed  for  longer  than  six  seconds  –  causes  restoring 
of manufacturer setting 

of communication protocol

, i.e. sets Modbus RTU communication 

protocol, device address sets to 01h and communication speed to 9600Bd - after button press 
there  is  “dEF”  message  blinking  at  LCD  display.  Six  seconds  later  message  “dEF”  stays 
shown, it means manufacturer setting 

of communication protocol

 is done. 

 

For  communication  with  device  T5410  which  measure  more  than  one  value,  there  is 

necessary to  add at the end of command, number of communication channel, where the measured 
value is mapped. Command for value reading is 

#AAx(CRC) cr

, where 

AA

 is device address, 

x

 is 

number of communication channel, 

CRC

 is check sum (can be used or not). 

 

Measured value 

Number of communication channel 

Temperature 

Atmospheric pressure 

 

Command  #AA(CRC) cr

  for  reading  all  measured  values  at  once  is  supported  for  multi-

channel devices since firmware version 02.60.  

 

Response: > (temperature)(relative humidity)(dew point temperature)(absolute humidity) 
  

(specific humidity)(mixing ratio)(specific enthalpy)(atmospheric pressure)cr 

 
 

ARION communication protocol - AMiT company 

The device supports communication protocol ARiON version 1.00. For more details see file 

Description of communication protocols of Txxxx series

” or  

www.amit.cz

. 

 
 

Communication with HWg Poseidon units 

 

Device  supports  communication  with  HWg-Poseidon  units.  For  communication  with  this 

unit set the device with setup software TSensor to communication protocol HWg–Poseidon and set 
correct  device  address.  This  communication  protocol  supports  read  temperature  at  °C,  relative 
humidity,  one  of  computed  value  (dew  point  temperature  or  absolute  humidity)  and  barometric 

Содержание T0310

Страница 1: ...nd improvement of their products Manufacturer reserves the right to make technical changes to the device without previous notice Misprints reserved Manufacturer is not responsible for damages caused b...

Страница 2: ...ware TSensor for setting of all device parameters recommended It is free to download at www cometsystem com It supports make the adjustment of the device too This procedure is described at file Calibr...

Страница 3: ...nection to RS232 interface For devices with RS485 output there is recommended to use shielded twisted copper cable maximal length 1200m The cable must be located at indoor rooms Nominal cable impedanc...

Страница 4: ...4 IE SNC T03 4 10 T5410 06 Dimensions T0410 T5410 Transmitter with RS485 Dimensions T0310L T0410L T5410L Connection see Appendix B...

Страница 5: ...IE SNC T03 4 10 T5410 06 5 Typical application wiring connection of terminals T0310 power from communication port T0310 power from external voltage source T0410 T5410 RS485...

Страница 6: ...tocols including examples of communication is available in individual document Description of communication protocols of Txxxx series which is free to download at www cometsystem com Note After switch...

Страница 7: ...R Serial number of device Lo 0x1036 4150 BCD BIN16 R Version of Firmware Hi 0x3001 12289 BCD BIN16 R Version of Firmware Lo 0x3002 12290 BCD BIN16 R Addition for device T5410 with atmospheric pressure...

Страница 8: ...pressed for longer than six seconds causes restoring of manufacturer setting of communication protocol i e sets Modbus RTU communication protocol device address sets to 01h and communication speed to...

Страница 9: ...it of allowed full scale range There is a reading Err1 on LCD display Value read from the device is 999 9 for pressure reading there is 999 9 hPa correct value This state occurs if measured temperatur...

Страница 10: ...ameters Ambient temperature RTD sensor Pt1000 3850ppm Measuring range 30 to 80 C Resolution of display 0 1 C Accuracy 0 4 C Atmospheric pressure T5410 only Unit hPa mBar PSI mmHg inHg inH2O oz in2 kPa...

Страница 11: ...ions other than specified in technical parameters Devices are not designed for locations with chemically aggressive environment Temperature sensor must not be exposed to direct contact with water or o...

Страница 12: ...the corresponding pins of CANON connector for more information see the operation manual for ELO E214 The ELO E06D converter is an optional accessory for connection of transmitter with RS485 interface...

Отзывы: