IE-SNC-T23(4)14-04 

 

  7 

Modbus registers of the device 

Variable 

Unit 

Address[hex]

X

 

Address[dec]

X

 

Format 

Size 

Status 

Address of device 

[-] 

0x2001 

8193 

Int 

BIN16 

R/W* 

Code of 
communication 
speed 

[-] 

0x2002 

8194 

Int 

BIN16 

R/W* 

Serial number of 
device Hi 

[-] 

0x1035 

4150 

BCD 

BIN16 

Serial number of 
device Lo 

[-] 

0x1036 

4151 

BCD 

BIN16 

Version of 
Firmware Hi 

[-] 

0x3001 

12289 

BCD 

BIN16 

Version of 
Firmware Lo 

[-] 

0x3002 

12290 

BCD 

BIN16 

Atmospheric 

pressure 

hPa  

0x0034 

52 

Int*10 

BIN16 

PSI  

Int*1000 

inHg 

Int*100 

mBar 

Int*10 

oz/in

2

 

Int*10 

mmHg 

Int*10 

inH

2

Int*10 

kPa  

Int*100 

 

Explanation: 

 

* depends on device setting (by User’s software) 

 

Int*10 register is in format integer*10 

 

register is designed only for reading 

 

W* 

register  is  designed  for  writing,  for  details  see  file  “

Description  of  communication 

protocols of Txxxx series

” 

 

X

 

register  addresses  are  indexed  from  zero  –  register  0x31  is  physically  sent  as  value 

0x30, 0x32 as 0x31 (zero based addressing). 

 

Note:

 In case there is a need for reading of measured values from the device with higher resolution 

than one decimal, measured values in device are stored also in „Float“ format, which is not directly 
compatible with IEEE754. 
 

Protocol compatible with Advantech-ADAM standard  

Control units communicate on master-slave principle in half-duplex operation. Only master 

can send requests and only addressed device responds. During sending request any of slave devices 
should  respond.  During  communication  data  is  transferred  in ASCII  format  (in  characters).  Each 
Byte  is  sent  as  two  ASCII  characters.  Device  supports  communication  speed  from  1200Bd  to 
115200Bd,  parameters  of  communication  link  are  1  start  bit  +  eight  bit  data  word  (LSB  first)  + 
1 stop bit, without parity. 

Jumper 

Jumper is located next to connection terminals. If communication protocol compatible with 

standard Advantech-ADAM is selected, its function is the following: 

 

If jumper during switching ON the power is closed, device always communicates 
with following parameters regardless stored setting in the device: 
communication speed 9600 Bd, without check sum, device address 00h 

Содержание T2314

Страница 1: ...cturer reserves the right to make technical changes to the device without previous notice Misprints reserved Manufacturer is not responsible for damages caused by using the device in conflict with thi...

Страница 2: ...www cometsystem com It supports make the adjustment of the device too This procedure is described at file Calibration manual pdf which is installed commonly with the software Change of some parameter...

Страница 3: ...r side The cable should be led in one line i e NOT to tree or star Termination resistor should be located at the end For short distance other topology is allowed Terminate the network by a termination...

Страница 4: ...4 IE SNC T23 4 14 04 Dimensions T2314 Transmitter with RS232 Dimensions T2414 Transmitter with RS485 Dimensions T2314L T2414L Connection see Appendix B...

Страница 5: ...IE SNC T23 4 14 04 5 Typical application wiring connection of terminals T2314 power from communication port T2314 power from external voltage source T2414 RS485...

Страница 6: ...bit registers Read Holding Registers 04 0x04 Reading of 16 bit input gates Read Input Registers 16 0x10 Setting of more 16 bit registers Write Multiple Registers Jumper and button Jumper and button ar...

Страница 7: ...g Note In case there is a need for reading of measured values from the device with higher resolution than one decimal measured values in device are stored also in Float format which is not directly co...

Страница 8: ...ce with setup software TSensor to communication protocol HWg Poseidon and set correct device address This communication protocol supports read temperature at C relative humidity one of computed value...

Страница 9: ...sure or error state of value If selected pressure unit is mBar or oz in2 or mmHg or inH2O or kPa there is shown only value number without corresponding pressure unit symbol 3 near by left display marg...

Страница 10: ...allowed manipulation It is not allowed to operate the device under conditions other than specified in technical parameters Devices are not designed for locations with chemically aggressive environmen...

Страница 11: ...connector marked RS232 directly to the PC connect power to connector marked RS485 Power voltage 6V DC from an external acdc adapter connect to pin 9 0V connect to pin 5 Also mutually connect pin 2 and...