Battery replacement 

Low  battery  is  indicated  on  the  display  by  blinking  of  reading  "BAT".  It  can  be  displayed 

permanently, if battery voltage is too low. Replace the battery for new one. Applied is Lithium battery 
3,6 V, size AA. Battery is located under logger lid. 

Warning:

  near  battery  fragile  glass  reed  contact  is  located  –  be  careful  not  to  damage  it.  Be 

careful in battery replacement! 

 

Replacement procedure: 

 

switch off the logger by the PC program or by magnet (if low battery allows) 

 

unscrew four corner screws and remove the lid 

 

remove old battery by pulling the glued tape 

 

insert new battery 

respecting the correct polarity

 (see s and – near battery holder). If 

you connect  new battery up to  30 s, all logger settings remain unchanged.  In the opposite case 
check by means of PC program all settings, especially the real time clock in logger. 

Attention, 

inserted battery with incorrect polarity causes logger damage!

 

 

put the lid back again and screw four screws. Be careful to have the rubber sealing properly in the 
slot and tighten the screws properly to ensure the water resistance of the instrument. 

 

connect logger to the computer and 

write to it the information on battery replacement

. This 

step is necessary to evaluate properly the remaining battery capacity. 

 

Old battery or logger itself (after its life) is necessary to liquidate ecologically! 
 
 

 

 

Instrument passed through tests of electromagnetic compatibility (EMC): 

 

Device conforms in accordance with EN 61326-1 these norms: 

 

radiation:  

EN 55011 Class B 

 

immunity:  

EN 61000-4-2 (levels 4/8 kV, Class A) 

 

 

 

EN 61000-4-3 (intensity of electric field 3 V/m, Class A) 

 

 

 

EN 61000-4-4 (levels 1/0.5 kV, Class A) 

 

 

 

EN 61000-4-6 (intensity of electric field 3 V/m, Class A) 

 
 
 

Technical support and service 

Technical support and service is provided by distributor. Contact is included in warranty 

certificate. 
 

Содержание S5141

Страница 1: ...LOGGER S5141 Three channel 0 5 Vdc datalogger with internal temperature sensor and display Instruction Manual...

Страница 2: ...nction Also memory alarm mode is selectable when alarm is indicated permanently till alarm memory reset Alarm function is enabled or disabled for each channel individually Logging mode can be adjusted...

Страница 3: ...normal environment accordingly to appendix NM AE1 AN1 AR1 BE1 Operational position negligible Logger installation by self adhesive Dual Lock applied to clean flat surface Not allowed manipulation inco...

Страница 4: ...er automatic switching ON or disable this option select if record runs permanently or only if alarm is active if alarms are about to be applied set both limits for each measured quantity and enable al...

Страница 5: ...the display in usual operation logger switched ON After switching ON the logger all LCD symbols are displayed for checking the display Then actual date and time in logger is displayed for approximate...

Страница 6: ...bol is replaced by adjacent symbol hyphen It appears in case all measured values are inside of adjusted alarm limits and data logging therefore does not run Displayed symbol indicates logger is ON If...

Страница 7: ...lled in non cyclic logging mode logger is switched OFF and message MEMO FULL appears on the LCD It appears also if logger is operated with switched OFF display New initialization of logger can occur i...

Страница 8: ...ogger OFF by magnet The procedure is identical with the above procedure for switching ON If decimal point does not appear after 1 s it is necessary remove the magnet and repeat the procedure Reset of...

Страница 9: ...cially the real time clock in logger Attention inserted battery with incorrect polarity causes logger damage put the lid back again and screw four screws Be careful to have the rubber sealing properly...

Отзывы: