background image

 

  IE-LOG-S5141-01 

Connecting logger to the computer by means of communication adapter 

 
 
 
 
 
 
 
 
 

 
 
 
 

 

Notice:

 USB connector can be located also from the computer front side 

 
After connecting the logger to the computer reading of logger info is enabled by means of the 

PC software and also setting of the instrument accordingly to the user needs (menu 

Configuration / 

Setting of instrument parameters

). 

 

 Before logging start it is necessary: 

 

 

check or optionally set the logger real time clock 

 

select suitable logging interval 

 

select logging mode (cyclic or non-cyclic) 

 

for channels IN1 to IN3 to specify conversion constants to the measured voltage values can 
be displayed directly in real physical readings and enter the number of decimal places 

 

switch OFF inputs that are not be recorded (e.g. when the sensor is not connected) 

 

switch  ON  the  logger  (or  switch  OFF,  if  it  is  about  to  be  switched  ON  by  magnet  or 
automatically with delayed start) 

 

enable or disable the option to switch ON the logger by magnet 

 

enable or disable the option to switch OFF the logger by magnet 

 

enable or disable the option to reset memory of  MIN/MAX values by magnet 

 

set date and time of logger automatic switching ON or disable this option 

 

select if record runs permanently or only if alarm is active 

 

if alarms are about to be applied, set both limits for each measured quantity and enable alarm 

 

optionally enable permanent alarm indication (alarm with memory) 

 

switch ON or switch OFF display logger 

 

optionally switch ON displaying of MIN/MAX values on the LCD 

 

reset memory of MIN/MAX values (if needed) 

 

check free space in data memory, optionally erase data memory of the logger 

 

enter password if protection against unauthorized manipulation with the logger is necessary  

 

The measured values are stored at preset intervals, while the moments of saving are chosen (for 

greater clarity) so that took place in integral multiples of time.  For the interval of 15 minutes the 
values will be stored every 15th, 30th, 45th and the 00th minute and zero second to every hour, at an 
interval of 3 clocks to be recorded at 00.00, 03.00, 06.00 hour etc. It follows that after starting the 
recording the instrument waits for the nearest integer multiple of time and then store the first value. 
This should be taken into account when set the time of the automatic switching ON of the logger. 

The external  sensor can  be connected only  when the logger is  switched off, otherwise it may 

cause incorrect writing to MIN / MAX memory and to memory of the measured values. 

Содержание S5141

Страница 1: ...LOGGER S5141 Three channel 0 5 Vdc datalogger with internal temperature sensor and display Instruction Manual...

Страница 2: ...nction Also memory alarm mode is selectable when alarm is indicated permanently till alarm memory reset Alarm function is enabled or disabled for each channel individually Logging mode can be adjusted...

Страница 3: ...normal environment accordingly to appendix NM AE1 AN1 AR1 BE1 Operational position negligible Logger installation by self adhesive Dual Lock applied to clean flat surface Not allowed manipulation inco...

Страница 4: ...er automatic switching ON or disable this option select if record runs permanently or only if alarm is active if alarms are about to be applied set both limits for each measured quantity and enable al...

Страница 5: ...the display in usual operation logger switched ON After switching ON the logger all LCD symbols are displayed for checking the display Then actual date and time in logger is displayed for approximate...

Страница 6: ...bol is replaced by adjacent symbol hyphen It appears in case all measured values are inside of adjusted alarm limits and data logging therefore does not run Displayed symbol indicates logger is ON If...

Страница 7: ...lled in non cyclic logging mode logger is switched OFF and message MEMO FULL appears on the LCD It appears also if logger is operated with switched OFF display New initialization of logger can occur i...

Страница 8: ...ogger OFF by magnet The procedure is identical with the above procedure for switching ON If decimal point does not appear after 1 s it is necessary remove the magnet and repeat the procedure Reset of...