background image

 

  IE-LOG-S5141-01 

Instruction manual for use of the S5141 logger 

 

The  device  is  designed  to  measure  and  record  signals  from  up  to  three  external  sensors  with 

output 0 - 5 Vdc with the possibility to record simultaneously the ambient temperature measured by 
the built-in sensor. Sensor connection is realized via terminals. Measured values are displayed on the 
two-line LCD display and are stored in selectable time interval to internal nonvolatile memory. All 
logger settings and control are performed by means of personal computer and password protection is 
enabled. It is also possible to switch ON and switch OFF the logger by start/stop magnet (this function 
can be disabled from the PC). Also automatic switch ON at programmed day and time (maximum 
one month forward) is enabled. Start/stop magnet enables also to clear the minimum and maximum 
value memory. 

Minimum and maximum measured values can be displayed (display switches to actual measured 

values and min/max values automatically). It is also possible to operate logger with switched OFF 
display. Short display of actual measured values is enabled by means of magnet. 

Switched ON logger every 10 seconds (independently on logging interval) updates MIN/MAX 

memory, compares measured values of each channel with two adjustable limits for each channel and 
exceeding  of  limits  is  indicated  on  the  display  (alarm  function).  Also  memory  alarm  mode  is 
selectable, when alarm is indicated permanently till alarm memory reset. Alarm function is enabled 
or disabled for each channel individually.  

Logging mode can be adjusted as non-cyclic, when logging stops after filling the memory. In 

cyclic mode oldest stored values are overwritten by new. In addition logging mode can be selected 
when logging is active only if measured value is out of adjusted alarm limits. 

Stored  values  can  be  transferred  from  logger  memory  to  the  PC  by  means  of  communication 

adapter. Communication adapter can be connected to the logger permanently  – data logging is not 
interrupted even if data download appears.  

Logger evaluates minimum battery voltage and its drop below allowed limit is indicated on the 

display. At the same time value of remaining battery capacity is available by means of the PC program 
and appears on the logger LCD in % (every time after switch ON). 

 
 
 
 
 
 
 
Technical parameters: 

 

Measuring parameters: 
  Voltage inputs (IN1, IN2 and IN3): 

Voltage measuring range: 0 to 5 V dc 
Effective resolution: 16 bits 
Accuracy: ±0.2 % FS 
Input resistance: approximately 100k ohms 

Ambient temperature (RTD sensor Pt1000/3850ppm): 

 

Measuring range: -30 to +70 °C 

   

 

Resolution: 0.1 °C 

   

 

Accuracy: ± 0.4 °C 
Response time: t63 < 6 min, t90 < 15 min (air flow approximately 1 m/s, 

temperature step 20 °C) 

Connection of input signals: shielded cable max. length of 10 m with terminal WAGO 734, 
   

 

 

(max. wires cross-section 1,5 mm

Measuring interval, alarm evaluation, MIN/MAX memory update and display refresh: 10 s 
Logging interval to memory: 10 s to 24 h (20 steps) 

Содержание S5141

Страница 1: ...LOGGER S5141 Three channel 0 5 Vdc datalogger with internal temperature sensor and display Instruction Manual...

Страница 2: ...nction Also memory alarm mode is selectable when alarm is indicated permanently till alarm memory reset Alarm function is enabled or disabled for each channel individually Logging mode can be adjusted...

Страница 3: ...normal environment accordingly to appendix NM AE1 AN1 AR1 BE1 Operational position negligible Logger installation by self adhesive Dual Lock applied to clean flat surface Not allowed manipulation inco...

Страница 4: ...er automatic switching ON or disable this option select if record runs permanently or only if alarm is active if alarms are about to be applied set both limits for each measured quantity and enable al...

Страница 5: ...the display in usual operation logger switched ON After switching ON the logger all LCD symbols are displayed for checking the display Then actual date and time in logger is displayed for approximate...

Страница 6: ...bol is replaced by adjacent symbol hyphen It appears in case all measured values are inside of adjusted alarm limits and data logging therefore does not run Displayed symbol indicates logger is ON If...

Страница 7: ...lled in non cyclic logging mode logger is switched OFF and message MEMO FULL appears on the LCD It appears also if logger is operated with switched OFF display New initialization of logger can occur i...

Страница 8: ...ogger OFF by magnet The procedure is identical with the above procedure for switching ON If decimal point does not appear after 1 s it is necessary remove the magnet and repeat the procedure Reset of...

Страница 9: ...cially the real time clock in logger Attention inserted battery with incorrect polarity causes logger damage put the lid back again and screw four screws Be careful to have the rubber sealing properly...

Отзывы: