![Comet System H5021 Скачать руководство пользователя страница 18](http://html1.mh-extra.com/html/comet-system/h5021/h5021_instruction-manual_2640755018.webp)
IE-HGS-H5(6)0XX-05
18
9.1.6. Value computed from measured relative humidity and temperature
Dew point temperature
Range:
-60 to +80 °C (-76 to 176 °F)
Accuracy:
±1,5°C (±2,7°F) at ambient temperature T<25°C (77°F) and relative
humidity RH >30%, for more details see graphs below
Absolute humidity
Range:
0 to 400 g/m
3
Accuracy:
±1,5 g/m
3
at ambient temperature T < 25°C (104 °F), for more details see graph
below