IE-HGS-H5(6)0XX-05
12
CO
2
measurement mode selection
This item provides to choice if measured value is displayed and read in “SLOW” or “FAST”
mode. “FAST” mode provides last measured value without averaging; this means that no software
filter is used for calculating the output value. “SLOW” mode provides value averaged from last 11
measured values.
averaged value „SLOW“ mode
current value without averaging „FAST“ mode
CO
2
zone level indication setup
Whole measuring range can be divided into three areas. Each active
area there is indicated by green, yellow or red LED depends on actual
measured value. It is done by two boundary limits setting. „YELL“
limit sets boundary between green and yellow indication. It means that
if actual measured value is lower than „YELL“ limit, there is green
LED active. If actual measured value is higher than „YELL“ limit, then
there is yellow LED active. Yellow LED lights until actual measured value exceeds „RED“ limit. If
actual value is higher than „RED“ limit, red LED is active. If you don’t want to use this indication
then set “YELL” limit to 0.
.
„YELLOW“ limit setting
„RED“ limit setting
Show temperature at LCD display
This item provides to choice if actual measured temperature may be displayed at LCD display or
not. It is possible to change it after „
SET
“ key pressing.
show value on LCD
don’t show value on LCD
Show relative humidity at LCD display
This item provides to choice if actual measured relative humidity may be displayed at LCD
display or not. It is possible to change it after „
SET
“ key pressing.