INSTRUCTION MANUAL
CLCI-100
BULK CURRENT INJECTION PROBE
with
FCLC-100
CALIBRATION FIXTURE
1912 1 E l T o ro R d
●
Si l vera do, Ca lifo rni a 9 26 76
●
(9 49) 459 -96 00
●
co m-p o wer. com
REV051917
Method A:
For method A, once the appropriate amplitude is
measured during calibration process, the signal
generator output level is recorded with respect to the
test frequency.
During the test, the signal generator output level at any
given test frequency is equal to the respective signal
generator output level recorded during the calibration.
Method B:
For method B, once the appropriate amplitude is
measured during the calibration process, the forward
power of the test generator is recorded. The forward
power is measured with an RF power meter and power
sensor, typically connected to the forward port of a
directional coupler installed in series with the RF output
of the power amplifier.
During the test, the signal generator output level is
adjusted so that the forward power measured on the
power meter at any given frequency is equal to the
respective forward power level recorded during the
calibration.
In the following sections, instructions are included for both methods:
Instructions common for both METHOD A and METHOD B are shown in
BLACK text.
Any instructions which apply only for METHOD A are shown in GREEN text.
Any instructions which apply only for METHOD B are shown in BLUE text.