4.0 Using the CLCI-100 Probe with the FCLC-100 Fixture
Calibration fixtures provide a means by which current probes, including bulk current
injection (BCI) probes, can be calibrated to determine the transfer impedance,
insertion/coupling loss. Fixtures are also used to establish, or calibrate, power drive levels
into 50
Ω
or 150
Ω
systems for various conducted susceptibility/immunity tests performed
using BCI probes. These applications are discussed in more detail in the following
sections.
The FCLC-100 Fixture is specifically designed for use with the CLCI-100 BCI probe. The
intent of the fixture is to maintain the coaxial structure of the transmission line, while
allowing the probe to be installed around the center conductor of the coaxial line.
4.1 Installation of CLCI-100 Probe into FCLC-100 Fixture
Illustrated in Figure 5 is the procedure to be followed for installing the CLCI-100 BCI
Probe into the FCLC Calibration Fixture.
FIGURE 5 -
Installation of CLCI-100 Probe into FCLC Calibration Fixture
STEP 1
UNSCREW SIX (6)
THUMBSCREWS
ON TOP OF
FIXTURE AND
REMOVE
TOP PLATE
STEP 2
OPEN CLAMP
AS SHOWN ABOVE
AND LOWER
CLAMP INTO
CENTER OF
FIXTURE
STEP 3
ROTATE CLAMP
APPROX. 90
DEGREES, SO THAT
PROBE’S RF
CONNECTOR FACES
OUTWARD ON EITHER
OPEN SIDE OF
THE FIXTURE
STEP 4
CLOSE PROBE
AROUND FIXTURE
CENTER
CONDUCTOR,
HOOK BOTH
SPRING-LOADED
CLASPS ONTO
RESPECTIVE
BRACKETS AND
LOCK CLASPS
STEP 5
REPLACE TOP
PLATE OF
TEST
FIXTURE
STEP 6
REPLACE AND
TIGHTEN THE SIX (6)
THUMBSCREWS