Page 26 of 42
INSTRUCTION MANUAL
ARI-6000-100W
RF Power Amplifier
SECTI ON 5
-
OPERATION
1912 1 E l T o ro R d
●
Si l vera do, Ca lifo rni a 9 26 76
●
(9 49) 459 -96 00
●
co m-p o wer. com
REV011718
Item
Description
Comments
Startup Option:
Initial Gain
Gain: XX.X%
Where x = 00.0 to 100.0
Allows choosing between 0%
attenuation to 100% attenuation at
startup. Where 0% is maximum, and
100% is minimum.
Startup Option:
RS-232
On or
Off
Enables or disables the Front Panel
Controller’s onboard RS-232 serial port.
Startup Option:
*RS-422
On or
Off
Enables or disables the Front Panel
Controller’s onboard RS-422 serial port.
Startup Option:
GPIB Port
On or
Off
Enables or disable the Front Panel
Controller’s onboard GPIB port.
Startup Option:
Ethernet Port
On or
Off
Enables or disable the Front Panel
Controller’s onboard Ethernet port.
VSWR Alarm:
VSWR Alarm: X
Where
X = 0 to 20
The VSWR Alarm Threshold represents
the difference between the forward
power and the reflected power (in
dB). A higher value here represents a
configuration that is more sensitive to
VSWR.
Restore Defaults
Yes/No
This allows the user to restore ALL
settings and calibration parameters to
those set at the factory.
Display Brightness
25%, 50%, 75%, 100%
Allows for setting of the brightness of
the vacuum-fluorescent display.
GPIB Termination
LF Term. On/Off
Adds an extra Line Feed (If On) to the
GPIB.
Telnet Timeout
Time: X Min
Telnet idle-period timeout. X = 0 to 100
minutes, where 0 = infinite (no timeout)
Exit
Save? (Y/N)
Exits User Configuration mode and
returns the display to Run Mode. The
user is given the choice to save
options. Other than as mentioned
above, choosing “No” will result in all
changes being lost, and the previous
settings will remain in EEPROM.
*= May not be displayed if Option was not purchased.