Rinse the filter under running water and dry it before re-installing. 

g.

 

Fill clean reservoir tank with distilled water using either top or front manual fill  

4.

 

Clean and fill the Venturi reservoir  

NOTE: If you did not drain your Venturi water reservoir, please start at Step 4a. If you drained 
your reservoir and waste bottle prior to closing your office, please start at Step 4c. 

a.

 

Open the unit door to connect the drain tube to the Venturi reservoir’s fron

t draining 

port. Drain the reservoir completely.  

b.

 

Empty the container 

c.

 

Using a soft bristle brush, scrub the corners and loosen any 
deposits.  

d.

 

Wipe the reservoir’s surfaces using OPT

IM

 wipes or a clean, 

lint-

free cloth dampened with water. Rinse the reservoir’s 

surfaces with distilled water and drain it using the front drain tube 

e.

 

Remove the reservoir filter by pulling up on the filter’s tab. 

 

f.

 

Rinse the filter under running water and dry it before re-installing. 

g.

 

When filling the Venturi reservoir after a period 
of non-use, you may need to add an additional 
0.5 US gal / 2 L of water. This will fill the steam 

cooling system’s condenser. To do this, simply fill 

the Venturi reservoir to the maximum fill line and 
wait as the Venturi reservoir water fills the 
condenser, then top up the Venturi reservoir to 
the maximum fill line again. 

5.

 

Run one empty Hollow Unwrapped cycle 

 

 

 

Note: if using an external water reservoir tank, ensure it is filled with water prior to starting the cycle 

Note: It is important to follow the recommendation of your local guidelines for the appropriate 
sterilization monitoring requirements after a period of non-use.  

For more information on performing machine maintenance, preparing unit for use and troubleshooting 
tips, please refer to the STAT

CLAVE

 

Operator Manual

.  

 