Potential Issues
Issue/CF: 17, 73, 77
Cause
•
Initial Vacuum phase unsuccessful
•
Unable to reach vacuum
Solution
•
Check door seal for any obstructions
•
Check that the door seal is properly seated in the door channel
•
Call Service
Issue/CF: 45
Cause
•
Water Pump timeout
Solution
•
Clean Water Reservoir too low
•
Reservoir water level needs to be monitored and refilled after 2-3 cycles
Issue/CF: 44
Cause
•
Venturi reservoir is too full
Solution
•
Ensure Venturi reservoir is filled to the max line
•
Check for blocked exhaust tube
•
Check the Venturi reservoir overflow sensor (the float sensor on the back wall of reservoir) and
make sure its not stuck in the high position.