 Potential Issues 

Issue/CF: 17, 73, 77 

 

Cause 

 

Initial Vacuum phase unsuccessful 

 

Unable to reach vacuum 

 

 

Solution 

 

Check door seal for any obstructions 

 

Check that the door seal is properly seated in the door channel 

 

Call Service 

 
Issue/CF: 45 

 

Cause 

 

Water Pump timeout 
 

Solution 

 

Clean Water Reservoir too low 

 

Reservoir water level needs to be monitored and refilled after 2-3 cycles 

Issue/CF: 44 

 

Cause 

 

Venturi reservoir is too full

 

 

Solution 

 

Ensure Venturi reservoir is filled to the max line 

 

Check for blocked exhaust tube 

 

Check the Venturi reservoir overflow sensor (the float sensor on the back wall of reservoir) and 
make sure its not stuck in the high position. 

 

 

 

 

 

 

 

 