STAT

CLAVE

 Unit 

1.

 

Perform the following maintenance procedures so the unit is clean before reservoirs are filled and 
instruments are processed.  

a.

 

Clean and disinfect the external surfaces using OPT

IM

 wipes or a clean, lint-free cloth 

dampened with water and, if needed, a mild detergent.  

i.

 

Dry the surfaces and remove any residue before powering ON the unit. 

b.

 

Clean the Sterilization Chamber, Rack and Trays 

i.

 

Remove the sterilization trays and the rack from the chamber. (Pull the rack out 
to disengage it). 

ii.

 

Use a clean, lint-free cloth dampened with water to clean the chamber and the 
chamber flange. Wipe dry. 

iii.

 

Use a clean, lint-free cloth dampened with water to clean the STAT

CLAVE

 rack 

and trays. 

 

 

c.

 

Inspect the door seal to ensure it is seated properly 

d.

 

Inspect and clean the 3 chamber filters 

i.

 

Remove the 3 chamber filters by unscrewing them from the chamber. Clean them 
under running water. Clear the mesh of debris. (If the filter cannot be reused, 
replace it - 01-115478S). 

ii.

 

To reinstall a filter, screw the filter back into position. The grooves on the filter 
head should be flush with the chamber surface to enable proper drainage. 
Tighten by hand. 

2.

 

Power the unit on. 

3.

 

Clean and fill the clean water reservoir.  

NOTE: If you did not drain your clean water reservoir, please start at Step 3a. If you drained 
your reservoir prior to closing your office, please start at Step 3c. 

a.

 

Open the unit door to connect the drain tube to 

the CLEAN WATER reservoir’s front draining 

port. Drain the reservoir completely.  

b.

 

Empty the container. 

c.

 

Using a soft bristle brush, scrub the corners and 
loosen any deposits.  

d.

 

Wipe the reservoir’s surfaces using OPT

IM

 

wipes or a clean, lint-free cloth dampened with 

water. Rinse the reservoir’s surfaces with 

distilled water and drain it using the front drain 
tube 

e.

 

Remove t

he reservoir filter by pulling up on the filter’s tab. 

 