General Description

The SI60 and SI60D incubators are constructed 
from clear acrylic material for total visibility. The 
units are designed for easy access with a hinged 
front door panel. Each side panel has 2 x 10mm 
diameter plugged holes for the introduction of 
either gases or cables. The incubator does not 
have a base so that it can be placed directly over 
complete instruments (an optional base and 
shelving system are available as accessories). 

The interior temperature is controllable up to 
60ºC; forced air circulation and electronic tem-
perature control ensure accurately maintained 
conditions. Additionally there is a temperature 
safety cut-out set at 72ºC.

The SI60 has an analogue controller while the 
SI60D has a digital display system. In addition to 
the main temperature selector there is a push 
switch temperature re-set which maintains the 
incubator at 37ºC.

Controls

Refer to Figures 1 & 2 to identify the controls.

Mains On/Off:

 This is a rocker-type switch. 

Pressing the top switches the power on and 
pressing the bottom switches the power off. The 
switch is illuminated when the unit is on.

Pre-set Temperature Switch:

 This is a 

rocker-type switch. Pressing the top switches the 
incubator to the pre-set temperature (37ºC). 
Pressing the bottom switches the incubator to 
the variable setting.

Load Indicator:

 An amber LED illuminates when 

the heater is in operation. This light remains on 
while the incubator is reaching the set 
temperature and flashes when the set 
temperature is reached.

Fault Indicator:

 This is a red warning light. 

When lit it indicates a fault has occurred and the 
temperature safety cut-off has been initiated (set 
at 72ºC).

Set Switch:

 This is a plain colour rocker switch. 

Pressing and holding the switch down displays 
the set temperature.

Temperature Control:

 This knob is used to set 

the temperature of the incubator in conjunction 
with the press to set switch.

Digital L.C.D.:

 This is used to display both the 

actual and set temperature of the incubator in 
ºC. The range is a5ºC - 60ºC with a 
resolution of 0.1ºC (SI60D only)

Figure 1: SI60 controls

Figure 2: SI60D controls

Temperature 
Control

Load
Indicator

Mains
On/Off

Fault

Indicator

Set

Switch

Pre-set

Temperature

Digital
Display

Temperature 
Control

Load
Indicator

Mains
On/Off

Fault

Indicator

Pre-set

Temperature

Содержание Stuart SI60

Страница 1: ...Incubator SI60 and SI60D Instruction Manual Version 1 2...

Страница 2: ...U K plug and replace the plug with a suitable alternative This involves cutting off the moulded plug preparing the cable and connecting to the re wireable plug in accordance with its instructions THI...

Страница 3: ...ower off The switch is illuminated when the unit is on Pre set Temperature Switch This is a rocker type switch Pressing the top switches the incubator to the pre set temperature 37 C Pressing the bott...

Страница 4: ...d back panels 7 Squeeze some silicon compound into the grooves on the back and top panel so when assembled a good seal is formed 8 Insert two side panels into back panel Then place top panel into posi...

Страница 5: ...internal repairs please contact the Service Department quoting both the model number and serial number Email cpservice coleparmer com Tel 44 0 1785 810475 Only spare parts supplied or specified by Co...

Страница 6: ...C to 60 C Temperature fluctuation at 37 C 0 1 C Temperature variation between shelves 0 3 C Nominal volume 60 litres Internal dimensions w x d x h 450 x 380 x 380mm External Dimensions w x d x h 600 x...

Страница 7: ...logue numbers The object of the declaration described above is in conformity with the relevant Union Harmonisation Legislation References to the relevant harmonised standards used or references to the...

Страница 8: ...Cole Parmer Ltd Beacon Road Stone Staffordshire ST15 0SA United Kingdom Tel 44 0 1785 812121 Email cpinfo coleparmer com Web www stuart equipment com...

Отзывы: