24
Page 28 of 41
M7886 issue 8.41
Push fit the inerting gas supply pipeline onto the inerting gas inlet port.
Turn on the gas and check for leaks.
Turn on the gas tap for each inerting cap position.
Your Omni Reaction Station is now ready for operation as described at the beginning of this
section.
7.1
1
. Cooling Plug (Optional) (10 Way - ATS1011
4)
The illustration below shows the main components of the cooling plug.
Figure 44: Cooling Plug
1
2
Coolant in and return quick release ports
Plug.
The cooling plug accessor
y is
specific
for
the 10 way dry blocks.
The c
ooling plug can be
connected to the coolant supply either in or out of the dry block. The plug can be inserted and
removed from the dry block whilst the reaction vessels are in the Reaction Station.
The cooling plug has a machined
aluminum
body which acts as the heat exchange interface
between the dry block and the chilled coolant supply. The plug body slides snugly into the dry
block, providing a good thermal contact between the cooling plug and dry block.