15
Page 19 of 41
M7886 issue 8.41
Figure 16: Setting the Stir Speed (Image 2)
Press the
^
key to make the 10’s value selection. Press the
v
key if you wish to reduce the
selected stir speed value.
Press the
>
key to return to the front screen menu.
OR…….
If 10 seconds elapse without a key being pressed the display will return to the main display
menu.
Note:
There is no facility for setting the units stir speed value.
To start the stirring process press the stir key marked with the symbol:
To stop the stirring process press the stir key once more.
Note:
To show stirring is in operation the stir symbol will remain illuminated. The stir speed
shown on the LCD is the
true speed
of stir. Not the setting.
7.6.
Heat Setting and Activation
7.6.1.
Setting the temperature when using a dry block without an internal sensor.
From the main display menu screen press the
<
key to go into the ‘heat setting’ screen.
Figure 17: Setting the Power
Note:
The default setting will always display 00%. Previous power settings are not retained.
Press the
>
key to go into set power selection screen.
Figure 18: Making the % Power Selection
Press the
^
key to make the
%
power selection. Press the
v
key if you wish to reduce the
power selection.
Note:
The power settings are made in increments of 5
C. The power selection range is from 0
to 100%.
Press the
>
key to return to the front screen menu.
Note: There is no facility for setting the units stir speed value.
Содержание RS-200 Series
Страница 1: ...Cole Parmer RS 200 Series Omni Reaction Station Instruction Manual M7886 Version 8 5...
Страница 34: ...34 NOTES...