8.1. Adjustment-
L
A.C
/
L
T.C
AND LA.N/LT.N
Models
8. BASIC OPERATIONS
2
1
Mass
standard [g]
1
000
2 000
5 000
5 000
10 000
TT-1003
TT-2102
TT-4502
TT-6002
TT-6001
TT-10001
Balance
model
Mass
standard [g]
50
100
200
200
200
500
L
A-64
L
A-124
L
A-214
L
A-314
L
T-363
L
T-603
Balance
model
Prior adjustment prepare the right
mass standard. You are recommended
to use mass standard of class F1 or F2.
View the table and select the mass
standard you need.
External adjustment is carried out using an external
mass standard of the right accuracy and weight
value, which value depends on balance type and
capacity.
8.
2
. Adjustment-
L
A AND
L
T
Models
2
1
4
3
Load the weighing pan with the
required mass standard (refer to
table above). Press key
1 000
Press key to start the adjustment
Wait for the adjustment completion.
Press key to start the adjustment
Remove all items from weighing
pan. Press key
Balance prepares for adjustment.